3
C. Specification Chart
Parameters
Ranges
Resolution
Accu-
racy
Auto
Tempera-
ture Com-
pensation
Adjustable
Tempera-
ture Com-
pensation
Cond/TDS
Ratios
Prepro-
grammed
Adjust-
able
Cond/
TDS
Ratio
Factor
Conductivity
0-9999µS/cm
10-200mS/cm
in 5 autoranges
0.01(<100µS)
0.1(<1000µS)
1(<10mS)
0.01(<100mS)
0.1(<200mS)
±1% of
reading
0-71ºC
32-160ºF
0-9.99%/ºC
KCl, NaCl,
442™
0.20-
7.99
TDS
0-9999ppm
10-200ppt
in 5 autoranges
0.01(<100ppm)
0.1(<1000ppm)
1(<10ppt)
0.01(<100ppt)
0.1(<200ppt)
±1% of
reading
0-71ºC
32-160ºF
0-9.99%/ºC
Resistivity
10K
Ω
-30M
Ω
0.01(<100K
Ω
)
0.1(<1000K
Ω
)
0.1(>1M
Ω
)
±1% of
reading
0-71ºC
32-160ºF
0-9.99%/ºC
pH
0-14pH
0.01pH
±0.01pH
*
0-71ºC
32-160ºF
ORP
±999mV
1mV
±1mV
*
Free Chlorine
Equivalent
(FC
E
)
0.00-9.99ppm**
0.01ppm
±0.3ppm
<1.00ppm
*
±0.2ppm
≥1.00ppm
*
0-71ºC
32-160ºF
350≤ORPmV<725
and 0.0≤pH<9.9
725≤ORPmV<825
and 0.0≤pH<8.9
Alkalinity Titration
10-800ppm
0.1ppm
Hardness Titration
0-1710ppm
(0-100 grains)
0.1(<1000 ppm)
1(<1710 ppm)
0.1(<100 grains)
LSI Calculator
-10 to +10
0.01
Temperature
0-71ºC
32-160ºF
0.1ºC/ºF
±0.1ºC
*EM Susceptibility: When Measuring 10.0 pH Reference Solution
±0.2 pH in the presence of RF fields ≥ 3 V/m.
±0.37 pH in the presence of RF fields at 300 MHz (±30MHz)
**If either ORP or pH is outside the specified limits, the instrument will display "-Or-".
WARNING: These products can expose you to chemicals including
Di(2-ethylhexyl)phthalate (DEHP), which is known to the State of
California to cause cancer and birth defects or other reproductive harm.
For more information go to www.P65Warnings.ca.gov
D. Warranty/Service
The Myron L
®
Ultrameter
III
, excluding the pH/ORP sensor, has a Two
(2) Year Limited Warranty. The pH/ORP sensor has a Six (6) Month
Limited Warranty for materials and workmanship. If an instrument fails
to operate properly, see Troubleshooting Chart, pgs. 52-53. The battery,
pH/ORP sensor and cell extender are user-replaceable. For other
service, return the instrument prepaid to the Myron L
®
Company.
MYRON L
®
COMPANY
2450 Impala Drive
Carlsbad, CA 92010-7226 USA
+1-760-438-2021
E-Mail: [email protected]
www.myronl.com
If, in the opinion of the factory, failure was due to materials or
workmanship, repair or replacement will be made without charge. A
reasonable service charge will be made for diagnosis or repairs due to
normal wear, abuse or tampering. This warranty is limited to the repair or
replacement of the Ultrameter
III
only. The Myron L
®
Company assumes
no other responsibility or liability.
TM