44
5.
Press
or
to change the amount of time (see
Figure 26). Maximum time of 75 seconds is shown.
6.
Press
to accept the change (new time).
XV.
USER Mode CALIBRATION LINC™ FUNCTION
The Linc™ function allows easy calibration when in User mode and the
user does not have a user standard solution to calibrate the instrument.
This function will ensure more repeatable and accurate measurements
than many other calibration methods. It is recommended that this
function be used to provide the highest degree of confidence when the
Ultrameter
III
is used in User mode. When Linc is used, the User mode
is linked to another standard, i.e., if User and KCl are linked, a KCI
standard solution is used to calibrate the instrument. It is that simple.
A. Calibration of Ultrameter
III
for use in User mode
1.
Press
or
key.
2.
Calibrate the unit using a Standard Solution (ref.
CALIBRATION, pg. 26).
3.
Place the Ultrameter
III
in User mode (ref. SOLUTION
SELECTION, pg. 23).
4.
Verify/Set the calibration linc. (See below – Setting “User”
Mode Calibration “Linc”.)