Recalling test results and averages
The eter is able to store 500 test results with time and date automatically. If your meter has
stored 500 results, which is the maximum memory, the newest test result will replace the oldest
one. To recall your test memory,
the meter without a test strip inserted.
m
switch on
2.
To
press main button again, and you will enter the averages
the
button and the display shows "AVG" in the upper right corner and
the average value for the day. By continually pressing the right
button you will get to
7-day, 14-day, 30-day and 90-day
of your blood glucose values. You will see
the number of calculated days in the lower left corner and the number of calculated readings in
the lower right corner.
end review of tests in memory,
mode. Press
right hand
hand
the
1-day,
averages
1. Press the main button to switch from the time mode to the memory
First
you will see the
symbol in the upper left corner of the display. When you
press the
button, the latest result displayed. By pressing the right
button continually
test results will be shown in
order. You will
see the sequence number in the lower right corner and the year in the lower left
corner of the
followed by date and time of the measurement.
mode.
"mem"
right hand
is
hand
previous
date
display,
Caring for your meter
Maintenance
Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the
carrying case
If
is dropped or damaged, perform a quality control test
before doing a blood glucose test
Cleaning
Clean the outside of the meter with a damp cloth and mild soap
when not in use. the meter
,
to insure that the meter is still functioning correctly.
the meter
or detergent. Do not wet the test
strip port.
Always keep the metal contact points of the test strip entry point clean. If any dust or impurities
are present, please clean with a small soft brush, otherwise the meter may not work correctly
when you insert a test strip.
NOTE
The right hand button is for reviewing the tests in ascending order, and the left hand button
is for reviewing in descending order. The number "1" is the latest test result, while the
number "500" is the oldest test result.