9.
replace the cap on the control solution
bottle.
10. The control solution result appears. Compare your quality control test result
to the control solution range printed on the test strip vial label.
Clean the top of the cap and tightly
Performing a quality control test
CLOSE
1. Take one test strip from vial and recap immediately.
2. Insert the test strip with
into
test strip port.
3. While the blood drop symbol
press and hold the main
button for
3 seconds until the "CS" symbol appears.
4. You will see
"
" and "CS" symbols on the screen,
prompting you to apply
solution.
5.
Open the bottle and
the cap
the table.
6. Drip a drop of control solution on the top of the cap.
7. Gently touch the sample entry of the test strip
control solution on top of the cap.
indication symbol upfront and down
is flashing,
more than
flashing
the control
Before opening the cap, shake the bottle of control solution well.
place
upright on
onto the drop of
8. When you hear a beep (if volume is switched on) please wait for
the test result. You will see the countdown mode on the screen.
After 5 seconds, the test result will appear.
Performing a quality control test
9
PRECAUTION
- Your control solution
will not be
average
but
be recalled. The control solution test result will be shown
"CS" symbol on
the screen.
- Our
for
solution test is 15~40
- Before " " and "CS" appear, please do not
the control solution to the sample
entry on test strip because the meter is still
an internal check. If you do so, the
meter will show " " and "Error", accompanied by beeps (if volume is
on).
- Do not drip the control solution
The reagent on
be sucked into the bottle of control
solution and
cause the
. Doing this
contaminate the meter via the test strip port.
- Do not touch the
of the control solution bottle. If you have touched it,
please clean
with water.
test results
included in
reading calculations,
can still
with the
suggested temperature range
the control
℃
(59~104°F).
present
performing
switched
directly from the bottle onto the sample entry of the test
strip.
the test strip could
could
solution's degeneration
could
potentionally
nozzle
it carefully
OPEN
4
5
6
2
7
1
8