![mylife PURA X Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/mylife/pura-x/pura-x_user-manual_1862630017.webp)
If control solution results are out of range, your BGMS may not be working properly. Repeat the
quality control test. If your control solution results
do not use the
BGMS.
Ypsomed customer service (see page 48).
are still outside the range,
Contact your local
33
32
TM
Quality control test with mylife Control control solution
PRECAUTION
Each time you open a new bottle of control solution, write the
discard date on the label.
solution is good for 3
months after opening the bottle, or until the
date printed
on the label, whichever comes first.
The control
expiry
Possible reasons
- our control solution has expired or has been opened for more than 3 months.
expired or has been opened for more than 6 months.
- Your control solution is diluted.
has been left open
The test procedure was not performed
- Your control solution test has been performed out of normal temperature range (<10°C and
>40°C resp. <50°F and >104°F).
your control solution results are out of range:
Y
correctly.
Malfunction of the meter or the test strip.
- Your test strip has
- The cap of the test strip vial or the control solution
for a long time.
-
-
Quality control
TM
test with mylife Control control solution
Please use control solution tested with BGMS under control solution mode. If the test result is
within the control solution range printed on the strip vial label, the BGMS passes
control test. That means your BGMS is working correctly.
the quality
Example of control solution range
printed on your test strip vial label.
Control solution range mg/dL mmol/L
Low 31-52 1,7-2,9
Normal 83-113 4,6-6,3
High 237-321 13,2-17,8
Control solution range:
When should a quality control test be performed?
Whenever you want to check
your BGMS is working
Whenever you want to practice testing and check correct procedure.
whether or not
properly.
Example of
date:
expiry
Exp. date 2017-10
NOTE
Control solution test results do not represent your blood
glucose level.