MGS Installation and Operation Manual
22
1000-0085 Rev 0
Section 6. Functional Tests and Calibration
6.1. Introduction
To comply with the requirements of EN378 and the European F-GAS
regulation, sensors must be tested annually. However, local regulations may
specify the nature and frequency of this test.
CAUTION:
Check local regulations on calibration or testing
requirements.
CAUTION:
The MGS contains sensitive electronic components
that can be easily damaged. Do not touch nor disturb any of these
components
NOTE:
The MGS is calibrated at the factory. After installation, a
zero adjustment maybe required due to differences in
environmental conditions.
IMPORTANT:
If the MGS is exposed to a large leak it should be
tested to ensure correct functionality by electrically resetting the
zero setting and carrying out a bump test. See procedures below.
IMPORTANT:
Murco recommends annual checks and gas
calibration. Murco also recommends sensor replacement every 3
years or as required. Calibration frequency may be extended
based on application, but should never exceed 2 years.
IMPORTANT:
In applications where life safety is critical,
calibration should be done quarterly (every 3 months) or on a
more frequent basis. Murco is not responsible for setting safety
practices and policies. Safe work procedures including calibration
policies are best determined by company policy, industry
standards, and local codes.
Содержание MGS
Страница 10: ...MGS Installation and Operation Manual 10 1000 0085 Rev 0 Figure 2 EC or IR Sensor Components and Wiring ...
Страница 11: ...MGS Installation and Operation Manual 1000 0085 Rev 0 11 Figure 3 SC Sensor Components and Wiring ...
Страница 17: ...MGS Installation and Operation Manual 1000 0085 Rev 0 17 Figure 5 IP66 Housing with Splashguard ...