20
3.2.2. RESISTANCE
Range
400
W
4k
W
40k
W
400k
W
4M
W
40M
W
Resolution
0,1
W
1
W
10
W
100
W
1k
W
10k
W
Intrinsic
uncertainty
±(0.8% R ±5 pt)
3.2.3. DIODE AND CONTINUITY
Function
Measurement range
Resolution
Intrinsic uncertainty
3V
1mV
The display unit indicates the nearest
standardized diode voltage.
The audible signal indicates a resistance
<50
W
Open-circuit voltage: 1V
3.2.4. CAPACITY
Range
10 nF
100 nF
1 µF
10 µF
100 µF
1 mF
10 mF
Resolution
1 pF
10 pF
100 pF
1 nF
10 nF
100 nF
1 µF
Intrinsic
uncertainty
± (3% R ±5 pt)
3.2.5. CURRENT
DC range
400µA
4000µA
40mA
400mA
10A
Resolution
0,1µA
1µA
10µA
100µA
10mA
Intrinsic
uncertainty
±(1% R ±5 pt)
±(2% R ±10 pt)
For AC measurements, the frequency is between 40 and 400Hz and the signal is sinusoidal.
Calibre AC
400µA
4000µA
40mA
400mA
10A
Resolution
0,1µA
1µA
10µA
100µA
10mA
Intrinsic
uncertainty
±(1% R ±5 pt)
±(2% R ±10 pt)
For the 10A current, the measurement time must not exceed 30 seconds every 15 minutes
The
10A
terminal is protected by a fuse: FF 10A H 600V 10 kA.
The
+
terminal is protected by a fuse: FF 400mA H 600V 10 kA.