21
3.2.6. FREQUENCY
In the Hz% setting
Input voltage
≥
2 V
ac
. This voltage increases with the frequency to be measured.
Range
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz
Resolution
0,001 Hz
0,01 Hz
0,1 Hz
1 Hz
10 Hz
10 Hz
100 Hz
Intrinsic
uncertainty
± (1% R ± 5 pt)
In the A or V setting
Maximum input voltage 600 V
ac
.
Input voltage
≥
600m V
ac
. This voltage increases with the frequency to be measured.
Measurement range
100 Hz
1000 Hz
10 kHz
100 kHz
Resolution
0,01 Hz
0,1 Hz
1Hz
10Hz
Intrinsic uncertainty
±(1.5% R ±5 pt)
3.2.7. DUTY CYCLE
Measurement range
0.1-99.9%
Resolution
0.1%
Intrinsic uncertainty
±3%
In the Hz% setting
Frequency: 1Hz to 10 MHz.
Input voltage
≥
2 V
ac
. This voltage increases with the frequency to be measured.
In the A or V settings
Frequency: 40Hz to 100 kHz.
Input voltage
≥
600 mV
ac
. This voltage increases with the frequency to be measured.
3.2.8. TEMPERATURE
Measurement range
-20°C ... +1000°C
Resolution
1°C
Intrinsic uncertainty
±(3% R ±3 pt)