4.
Other Components:
Check all knobs, printed circuit boards, screws, fasteners, connections and terminals for
tightness and proper position. Remove dust with a soft brush and breath of air. Output
terminal connection tightness is particularly important. If they become loose, excessive
heating of the terminals and poor current output will result.
5.
Insulation:
Check wiring and other insulated components for burning, cracking or other damage.
IMPORTANT NOTE
Do not use lubricants or solvents of any kind in the test set except as specifically recommended.
SERVICE AND REPAIR ORDER INSTRUCTIONS
If factory service is required or desired, contact the factory for return instructions.
A Service & Repair Order (SRO) number will be assigned for proper handling of the unit when it
arrives at the factory.
If desired, a letter with the SRO number and instructions can be provided.
Provide the factory with model number, serial number, nature of the problem or service desired,
return address, your name, and where you can be reached should the factory need to contact you.
A purchase order number, cost limit, billing, and return shipping instructions may also be provided if
desired.
National Bureau of Standards traceable calibration and certification of two types is available, if
desired, at additional cost:
Class One
A certificate is provided verifying the traceability and calibration of the
equipment.
Class N
That which is required for nuclear power plants. A certificate of traceability
and calibration along with "as found" and "as left" data are provided.
If an estimate is requested, provide the name and contact information of the person with