Model CB-832 is rated at 1.75 kVA output and is equipped with three output terminals, each capable 

of supplying its rated current.  The current rating of these output terminals may be exceeded for 

short durations provided the voltage rating is sufficient to "push" the desired current through the 

device under test and the connecting test leads.  The overload capacity, represented by multiples of 

rated current, versus TIME ON and TIME OFF are given below. 

 

%Rated 

 

Maximum 

 

Minimum 

Current 

 

Time ON 

 

Time OFF

 

 

100 (1x) 

 

30 minutes 

 

30 minutes 

200 (2x) 

 

 3 minutes 

 

 8 minutes 

300 (3x) 

 

30 seconds 

 

 4 minutes 

400 (4x) 

 

 7 seconds 

 

 2 minutes 

 

Model CB-832 also can be used continuously for any reasonable length of time at 70.7% of the 

current rating of the output terminal. 

 

EXAMPLE: 

Consider the terminal rated at 500 amperes: 

 

a. 

500 amperes may be drawn at any voltage from 0-3.5 volts for 15 minutes provided 

the test set is subsequently de-energized for 30 minutes. 

 

b. 

353.5 amperes (70.7% of 500 amperes) may be continuously drawn at any voltage 

from 0-3.5 volts for any reasonable length of time. 

 

c. 

The following overload currents may be drawn at any voltage from zero to maximum 

volts  for  the  Time  ON  indicated,  followed  by  the  indicated  Time  OFF.    The 

maximum output voltage available when the current rating of the output terminal is 

exceeded (overloaded) will be less than the rated value due to the regulation of the 

transformer in the test set.  For example, when drawing 750 amperes from the 500 

ampere terminal, the maximum voltage available is approximately 2.9 volts. 

 

 

500 Ampere   

 

 Time ON 

 

 Time OFF

 

 750 amperes   

 

6 minutes 

 

13 minutes 

1000 amperes   

3 minutes 

 

 8 minutes 

 

 

It should be noted that because of the impedance of the device under test and the connecting test 

leads,  the  maximum  practical  test  current  available  from  the  terminal  rated  500  amperes  is 

Содержание CB-832

Страница 1: ...ON MANUAL For HIGH CURRENT TEST SET MODEL CB 832 It is essential that this instruction book be read thoroughly before putting the equipment in service Advanced Test Equipment Rentals www atecorp com 8...

Страница 2: ...ly reserves to itself all rights to such proprietary information as well as all rights under any such patent none of which is waived by the submission of this instruction manual to anyone The recipien...

Страница 3: ...6 D Overload Capacity 7 III TEST PROCEDURES 8 A Maintenance of Motor Overload Relays 8 1 Planned Maintenance Program 10 2 Time Delay 11 3 Instantaneous Element 13 B Maintenance of Molded Case Circuit...

Страница 4: ...complete working understanding of the CB 832 test set It is divided into five main sections SECTION I Gives a description of the test set Gives basic guidelines on the different applications performe...

Страница 5: ...essential that the technician properly and effectively ground earth the test set This is accomplished by connecting the input line cord to a properly grounded receptacle CAUTION The input voltage of t...

Страница 6: ...er output circuit of CB 832 against overloads CONTROL FUSE Protects instrumentation and control circuitry against overloads OUTPUT CONTROL Variable autotransformer provides continuous non stepped outp...

Страница 7: ...power remains energized but output current remains de energized and timer will not count AMMETER LED displays show the numeral one in the left most position when the meter is over ranged RANGE SWITCH...

Страница 8: ...not connect CONTACTS binding posts to an energized circuit STOP MODE Switch Selects either current actuated position to stop the current output and timer or either of the two external contact stop mod...

Страница 9: ...ome experimentation and will be easier as familiarity increases NOTE There is no relationship between the ammeter ranges and the ratings of the output terminal All ammeter ranges can be used in conjun...

Страница 10: ...00 amperes may be drawn at any voltage from 0 3 5 volts for 15 minutes provided the test set is subsequently de energized for 30 minutes b 353 5 amperes 70 7 of 500 amperes may be continuously drawn a...

Страница 11: ...ference to tables or charts supplied by the manufacturer of the overload relays and motors Whenever a motor trips out it is poor practice to uprate the overload relay indiscriminately the motor may ac...

Страница 12: ...nergy in the spring This interrupts the circuit of the holding coil in the motor starter and shuts down the motor The starter may be reset only after the temperature of the heater has cooled sufficien...

Страница 13: ...ossible for dirt or dust created by conditions in the plant to prevent parts of the relay from moving These same conditions can also prevent the proper dissipation of normal heating resulting in unnec...

Страница 14: ...load relay Connect other end of this lead to the appropriate output terminal see SELECTION OF OUTPUT TERMINAL 4 Connect test set to suitable single phase power supply 5 Turn test set on with POWER ON...

Страница 15: ...may decrease fall off during the test because the resistance or impedance of the test circuit increases as it heats up Rotate OUTPUT CONTROL knob clockwise to keep test current at desired value 15 Wh...

Страница 16: ...Open Contacts of overload relay to binding posts of test set labeled CONTACTS 9 Select appropriate timer STOP MODE 10 Select desired timer display mode and range 11 Rotate OUTPUT CONTROL knob clockwis...

Страница 17: ...e is sometimes referred to as the jogging or step method For nearly all applications this method works best however there is another method referred to as the run up method Instead of the OUTPUT MODE...

Страница 18: ...ual inspections and electrical tests A brief outline is given below 1 Clean All types of molded case circuit breakers should be externally cleaned so that the heat produced in normal operation can be...

Страница 19: ...ument displays should light 6 Use RANGE Switch to select ammeter range so test current will be near full scale and no less than 10 of full scale 7 Put ammeter DISPLAY MODE Switch in MEMORY position 8...

Страница 20: ...r impedance of the test circuit increases as it heats up Rotate OUTPUT CONTROL knob clockwise to keep test current at desired value 14 When circuit breaker trips timer stops and output is de energized...

Страница 21: ...sition 9 Select desired timer display mode and range 10 Rotate OUTPUT CONTROL knob clockwise and momentarily press OUTPUT MODE Switch in MOMENTARY release Observe current reading retained by ammeter I...

Страница 22: ...ings and brush es can cause premature wear down of the brush es and carbon buildup see SERVICE DATA GROUND FAULT ELEMENT 1 Set up CB 832 with a POWER ON OFF Switch in OFF position instrument displays...

Страница 23: ...OUTPUT CONTROL knob at maximum clockwise rotation return knob to zero and transfer output lead to the next higher voltage lesser current terminal Proceed with current adjustment as in Steps 10 and 11...

Страница 24: ...cloth can be dampened with an approved solvent that does not attack the finish or leave residue 2 Control Panel The control panel can be wiped clean with a soft dry cloth Do not wipe the meter lenses...

Страница 25: ...SRO number will be assigned for proper handling of the unit when it arrives at the factory If desired a letter with the SRO number and instructions can be provided Provide the factory with model numbe...

Страница 26: ...mps and shocks of shipping via common commercial carrier For example you may wish to reship your unit to Multi Amp for annual calibration certification OPERATIONAL ERRORS TROUBLESHOOTING Some operatio...

Страница 27: ...ts connector has come loose from the PCB d Triac failure 5 Same as above will not initiate in N O or N C Contacts mode a Faulty test leads b Incorrect test scheme for stop mode selected c Contacts bei...

Страница 28: ...quipment which has damage or cause of defect due to accident negligence improper operation faulty installation by the purchaser or improper service or repair by any person company or corporation not a...

Страница 29: ...Part 12470 Rev 2 3 25 99 26 SECTION V SCHEMATICS...

Отзывы: