Temposonics
®
R-Series
Brief Instructions
5.4 CANbus
CANbus LED status
Green
Red
Information
ON
OFF
Normal function
ON
ON
No magnet / wrong quantity of magnets
OFF
ON
Initialization error
Flashing
Flashing Operating voltage out of range
D54
Signal
M12 male connector
(A-coded)
Pin
Function
1
2
3
4
5
View on sensor
1
Shield
2
Not connected
3
Not connected
4
CAN_H
5
CAN_L
M12 female connector
(A-coded)
Pin
Function
1
2
3
4 5
View on sensor
1
Shield
2
Not connected
3
Not connected
4
CAN_H
5
CAN_L
Power supply
M8 male connector
Pin
Function
2
4
1
3
View on sensor
1
+24 VDC (−15 / +20 %)
2
Not connected
3
DC Ground (0 V)
4
Not connected
D55
power supply
M12 male connector
(A-coded)
Pin
Function
1
2
3
4
5
View on sensor
1
Shield
2
+24 VDC (−15 / +20 %)
3
DC Ground (0 V)
4
CAN_H
5
CAN_L
M12 female connector
(A-coded)
Pin
Function
1
2
3
4 5
View on sensor
1
Shield
2
+24 VDC (−15 / +20 %)
3
DC Ground (0 V)
4
CAN_H
5
CAN_L
D60
power supply
M16 male connector
Pin
Function
1
2
3
6 5
4
View on sensor
1
CAN_L
2
CAN_H
3
Not connected
4
Not connected
5
+24 VDC (−15 / +20 %)
6
DC Ground (0 V)
D62
power supply
M16 male connector
Pin
Function
1
2
3
6 5
4
View on sensor
1
CAN_L
2
CAN_H
3
Not connected
4
Not connected
5
+24 VDC (−15 / +20 %)
6
DC Ground (0 V)
M16 male connector
Pin
Function
1
2
3
6 5
4
View on sensor
1
CAN_L
2
CAN_H
3
Not connected
4
Not connected
5
+24 VDC (−15 / +20 %)
6
DC Ground (0 V)
P
XX
/ H
XX
power supply
Cable
Color Function
GY
CAN_L
PK
CAN_H
YE
Not connected
GN
Not connected
BN
+24 VDC (−15 / +20 %)
WH
DC Ground (0 V)