Maintenance
Maintenance of the DCPD consists of keeping the unit clean and occasionally cleaning the filter.
The unit can be wiped off with a clean, damp cloth and then wiped dry with a clean, lint-free cloth.
Clean the Air Filter
The filter is stainless steel mesh and intended to be cleaned rather than replaced. It is recommended to check
the filter every 6 months (or less in dirty environments) and clean the filter if necessary. Perform the following
procedure to clean the filter.
Air filter
Description
Item
Air in. The filter is behind the cover.
1
Air out.
2
1. Disconnect the power cord.
2. Remove the four screws securing the right fan cover.
3. Remove the filter and clean it. The filter can be cleaned by blowing off with compressed air or cleaned
with soapy water, rinsed, and allowed to dry completely.
4. Replace the cleaned, dry filter.
5. Replace the cover removed in step 2.
6. Connect the power cord.
34
DCPD Measurement System
Maintenance
Содержание DCPD
Страница 1: ...DCPD Measurement System 100 279 690 B be certain...
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 17: ...Introduction Topics DCPD Measurement System Introduction 18 DCPD Measurement System 17...
Страница 21: ...Installation Topics DCPD Installation 22 DCPD Configuration 24 DCPD Measurement System 21...
Страница 28: ......
Страница 29: ...Operation Topics DCPD Operation 30 Foil Specimen 31 DCPD Measurement System 29...
Страница 32: ......
Страница 33: ...Maintenance Topics Maintenance 34 Clean the Air Filter 34 DCPD Measurement System 33...
Страница 35: ...Troubleshooting Topics Troubleshooting 36 DCPD Measurement System 35...
Страница 36: ...Troubleshooting Reason Check Issue 36 DCPD Measurement System Troubleshooting...
Страница 37: ...100 279 690 B...
Страница 38: ......