background image

DCPD Operation

Operation during testing is primarily controlled through software and the Test Controller. Refer to your software
and controller documentation for additional information.

Manual Controls

Description

Item

Current Set: 10-turn control and Local/Remote selector switch.

1

Gain: X1/X10 selector switch.

2

LPF (low pass filter) 50HZ/300HZ selector switch. This switch should typically be set
to

3

Manual operation is somewhat empirical and depends on specimen size and material. It is suggested that
you initially use these controls on dummy specimens and monitor results during simulated testing to achieve
familiarity with these controls.

Current Set and Gain

In most cases, the current is controlled remotely (Local/Remote switch set to Remote). With the Local/Remote
switch set to Local, the current is controlled by the 10-turn potentiometer.

The current gain goes from zero Amps (dial setting 000) to about 20 to 25 amps (dial setting 1000). A dial
setting of 500 would be about 10 to 12 amps.

Bigger specimens need more current to measure specimen resistance (Potential Drop) effectively. The typical
current for a room temperature w =2.0 inches B=1.0 inches aluminum specimen is about 10 amps. Thinner
specimens need less, thicker specimen need more.

The typical gain for the aluminum specimen mentioned above us about 50,000 times gain. The preamplifier
is 5000 times gain. The Gain X1/X10 switch can add another 10 times gain on top of the 5000 times gain the
preamp provides to raise the gain to 50000 times gain. Some small (thin) specimens will not need the additional
10 times gain.

30

DCPD Measurement System

Operation

Содержание DCPD

Страница 1: ...DCPD Measurement System 100 279 690 B be certain...

Страница 2: ...ietary confidential and owned by MTS Systems Corporation and cannot be copied reproduced disassembled decompiled reverse engineered or distributed without express written consent of MTS Software Verif...

Страница 3: ...Conventions 9 Safety 13 General Safety Practices 14 Hazard Placard Placement 14 No User Serviceable Parts Within 15 Introduction 17 DCPD Measurement System Introduction 18 Installation 21 DCPD Instal...

Страница 4: ......

Страница 5: ...de 1 800 328 2255 toll free in U S 1 952 937 4000 outside U S Telephone Europe 800 81002 222 International toll free in Europe Outside the U S For technical support outside the United States contact y...

Страница 6: ...el if applicable What is the system configuration Know relevant computer information For a computer problem have the following information available Manufacturer s name and model number Operating soft...

Страница 7: ...rodynamic material test system Electromechanical material test system Hydromechanical material test system Vehicle test system Vehicle component test system Aero test system Be prepared to troubleshoo...

Страница 8: ...ittal Form to communicate problems with your software hardware manuals or service that are not resolved to your satisfaction through the technical support process The form includes check boxes that al...

Страница 9: ...er and application software manuals are typically included on the software CD distribution disc s Documentation Conventions The following paragraphs describe some of the conventions that are used in y...

Страница 10: ...ighlight easily overlooked information Recommended Recommended notes provide a suggested way to accomplish a task based on what MTS has found to be most effective Tip Tips provide helpful information...

Страница 11: ...ertext links displayed in a blue font All blue words in the body text along with all contents entries and index page numbers are hypertext links When you click a hypertext link the application jumps t...

Страница 12: ......

Страница 13: ...Safety Topics General Safety Practices 14 Hazard Placard Placement 14 No User Serviceable Parts Within 15 DCPD Measurement System 13...

Страница 14: ...the following known variables Facility variables facility variables include the structure atmosphere and utilities Unauthorized customer modifications to the equipment Operator experience and speciali...

Страница 15: ...No User Serviceable Parts Within Part 572402 07 No user serviceable parts within DCPD Measurement System 15 Safety...

Страница 16: ......

Страница 17: ...Introduction Topics DCPD Measurement System Introduction 18 DCPD Measurement System 17...

Страница 18: ...ut these conditioners do not have enough current or amplification for typical DCPD specimens Therefore the DCPD Measurement System electronics provide the current and amplify the resulting potential d...

Страница 19: ...m supports a single reference specimen and a single active specimen Connects between the DCPD box and the reference specimen and active specimen s One preamp is required for every two specimens Preamp...

Страница 20: ...f 15 24 cm 6 in for cable connections Space Requirements Note To maintain EMC compliance the controller must be installed in a location that does not exceed the EN 55011 emission standard Height 149 2...

Страница 21: ...Installation Topics DCPD Installation 22 DCPD Configuration 24 DCPD Measurement System 21...

Страница 22: ...nsure optimal performance MTS recommends ordering all system cables from MTS Cables should be installed so that they are protected from conditions that could damage the cable When you are finished dou...

Страница 23: ...able Connections Description Item FlexTest Controller 1 Preamp 2 Active Specimen 3 Reference Specimen when applicable 4 Connection to electrical main power 5 Electrical and earth grounds 6 DCPD Measur...

Страница 24: ...rence specimen to the DCPD unit The Channel 2 green cable carries the resistance value from the active specimen to the DCPD unit The Channel 1 yellow cable from the DCPD connector into the preamp and...

Страница 25: ...per Plug 4 Required to verify that the controller is ready to run Service Port 5 Four analog inputs for voltage current and switch status Station Builder Current Command Digital Outputs Off DCPD Measu...

Страница 26: ...Manual Command Digital Output On Reverse Off Manual Command Digital Output On Reverse On Change Gain in DUC Active and Reference Channel Typical Voltages 26 DCPD Measurement System Installation...

Страница 27: ...rrent Output shown is half of the full scale command which is 11 amps Digital Output Off Reverse Off With these output settings no current is going to the specimen Only thermocouple voltages are being...

Страница 28: ......

Страница 29: ...Operation Topics DCPD Operation 30 Foil Specimen 31 DCPD Measurement System 29...

Страница 30: ...otely Local Remote switch set to Remote With the Local Remote switch set to Local the current is controlled by the 10 turn potentiometer The current gain goes from zero Amps dial setting 000 to about...

Страница 31: ...LPF If additional amplification or filtering is required the signals can be further conditioned by the MTS controller if they are brought in on a DUC LPF The LPF or low pass filter is used to filter...

Страница 32: ......

Страница 33: ...Maintenance Topics Maintenance 34 Clean the Air Filter 34 DCPD Measurement System 33...

Страница 34: ...less in dirty environments and clean the filter if necessary Perform the following procedure to clean the filter Air filter Description Item Air in The filter is behind the cover 1 Air out 2 1 Disconn...

Страница 35: ...Troubleshooting Topics Troubleshooting 36 DCPD Measurement System 35...

Страница 36: ...Troubleshooting Reason Check Issue 36 DCPD Measurement System Troubleshooting...

Страница 37: ...100 279 690 B...

Страница 38: ......

Отзывы: