13
Modbus Interface Manual
LP Series
9.1.9 Save settings
Fig. 11: Save settings
User Configurable:
Read Settings from File:
Allows the end user to upload factory
parameters from a backup file to the LP Dashboard. This task is
usually performed from a saved backup file or the original backup file
maintained by MTS.
Write Setting to a File:
Allows the end user to download a backup file
of factory parameters from the LP Dashboard to a PC. This task is
usually performed after Read Settings from Gauge. Note – wait until
all settings have changed from Red to White before writing as the color
change signals that the settings have been updated.
Write Settings to Gauge:
Allows the end user to program the level
transmitter with the factory parameters displayed on the LP Dashboard.
This task is usually performed after Read Settings from File.
Read Settings from Gauge:
Allows the end user to update all of the
factory parameters displayed on the screen. All settings will turn Red
and then will turn white as they are updated.
Note:
A copy of the backup file is maintained by MTS including all factory
parameters as the level transmitter was originally setup after
completing testing and calibration at the MTS factory. MTS can
provide a copy of the backup file upon request based off of the serial
number of the level transmitter. Contact MTS Technical Support for
assistance.
9.2 Programming the display
The display diagram is shown in section 6.2. The display menu structure
is shown in section 6.3. This section of 9.3 explains the details of the
programming available under the different sections of the display. The
factory password for entering the display is
27513
.
9.2.1 Data from Device
Display
Allows the user to select if the display shows Length Units or Volume
Units.
Units
Allows the user to change the selected Length Units, Volume Units and/or
Temperature Units.
Address
Allows the user to change the address of the level transmitter. Default
address is 247.
Signal Strength
Allows the user to view the strength of the return signal for the product
float (Prod Trig Lvl), interface float (Int Trig Lvl), and roof level (Roof Trig
Lvl). If the Interface float and or the roof level are not active no signal
can be viewed.
9.2.2 Calibrate
Product Level
Allows the user to change the level in engineering units for calibration.
The user should use the Current Level selection and enter the current
position of the float. The user is advised not to use the Offset feature
without help from Technical Support.
Interface Level
Allows the user to change the level in engineering units for calibration.
The user should use the Current Level selection and enter the current
position of the float. The user is advised not to use the Offset feature
without help from Technical Support.
Roof Level
Allows the user to change the level in engineering units for calibration.
The user should use the Current Level selection and enter the current
position of the roof. The user is advised not to use the Offset feature
without help from Technical Support.
9.2.3 Factory
Settings
Menu section that contains factory parameters. Do not edit these
parameters without talking to Technical Support.
Gradient
The gradient is a calibration factor that is unique for each transmitter.
Typical values are between 8.9 and 9.2 µs/in.
Serial Number
The Serial Number is the unique identifier for the unit from MTS and
should not be changed. The serial number is used for tracking and
determining spare parts.