![MTI 526 Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/mti/526/526_user-manual_1852292021.webp)
User Manual | MTI 527, 527W, &526
17
Cleaning and Disinfection
Upholstery
Upholstery should be cleaned weekly with a mild
liquid soap and warm water mixture. Afterwards, wipe
upholstery with clean damp cloth.
MTI’s standard policy on disinfecting MTI medical
equipment is to use a 10% sodium hypochlorite
(bleach) and 90% water solution. If stronger or other
disinfecting and cleaning solutions are used, the
surface is at risk to be damaged or possibly cause
discoloring, cracking, or other types of damage to the
surface. See current CDC Guideline for Disinfection &
Sterilization in Healthcare Facilities.
MTI understands that many customers prefer to use a
pre-mixed disinfecting solution in a spray or wipe. It is
MTI’s recommendation not to use sprays because over
time the overspray will cause damage to the surfaces
if not immediately wiped up. MTI recommends hand
towels or wipes to avoid the possibility of overspray.
Once the approved contact time has been obtained,
remove and dry excess liquid remaining on the
surface.
See “MTI Cleaning Solution Recommendation Letter”
for more details on disinfecting solutions which have
been tested on MTI’s products. It is available for
download at http://www.mti.net/downloads.aspx in
the Product Support Bulletins folder, found under the
Resources tab on our website. In the U.S. & Canada,
call (888) 520-4998, or International, call (801) 875-
4998 to request a copy.
Painted and Plastic Surfaces
All painted metal parts have been electrostatically
painted and the paint is baked on for durability. These
surfaces should be wiped down with a quality grade
cleaner and a soft cloth or a damp cloth with mild
soap once a week.
Preventative Maintenance
Periodically inspect the following areas:
1. All fasteners and pins should be in place and
tightened securely.
2. Lubricate pins and moving joints as needed with
lithium grease.
3. Power cord should be free of cuts or other visible
damage.
4. All mechanical functions should operate properly.
Have an authorized service technician inspect your
equipment once a year.
Fuse Replacement Instructions
1. Unplug the chair.
2. Open the fuse drawer below the female
connection for the power cord.
3. Remove the burnt fuse(s). (Needle nose pliers or
hemostats may be needed.)
4. Insert new fuse(s) with same type of fuse
removed in step 3. (See figure 1 below) See
”SPECIFICATIONS” on page 22 for fuse type.
5. Close the fuse drawer.
6. Reinsert the power cord into the base of the chair
and secure it with the retainer clip.
Figure 1
Содержание 526