Po ly Gar d ® 2 M SC 2  –  Dat a Sh eet  an d   Us er  M an u al  

Specifications subject to change without notice 

PolyGard® is a registered trademark of MSR-Electronic GmbH 

 

FUNCTIONAL TEST (FOR INITIAL OPERATION AND MAINTENANCE) 

The functional test should be carried out during each service, but at least once a year. 

ZERO-POINT TEST WITH FRESH OUTDOOR AIR 

Due to the employed optical measurement principle, there won’t be any deterioration caused by ageing or chemical 
processes in the sensor. However, a zero offset may occur. 

Using the Service Tool STL 06 you can read it out and document it in form of a measured value. 

TRIP TEST WITH REFERENCE GAS 

The sensor is gassed with reference gas CO

2

, concentration, 2-3 % vol, (for this you need a gas bottle with pressure 

regulator and a calibration adapter). 

In doing so, the set alarm thresholds are exceeded, and all output functions are activated. It is necessary to check if the 
connected output functions are working correctly (e.g. the horn sounds, the fan switches on, devices shut down). By 
pressing the push-button on the horn, the horn acknowledgment must be checked. After removal of the reference gas, 
all outputs must automatically return to its initial position. 

Other than the simple functional testing, it is also possible to perform a functional test by means of calibration. For 
further information, please refer to the User Manual. 

CALIBRATION 

New Sensor Cartridges SC2 are always delivered factory-calibrated by MSR-E. This is documented by the calibration label 
indicating date and calibration gas. A repeated calibration is not necessary during commissioning if the device is still in its 
original packaging (air-tight protection by the red protective cap) and the calibration doesn’t date back more than 12 
months for CO

2

 sensors and 3 months for all other gases. 

Due to high operational stability of the sensor, and the intrinsically safe design of the gas detector, the gas detector 
doesn’t need an annual calibration. 

After an operating period of 5 years, only the sensor cartridge must be recalibrated & adjusted by the manufacturer or 
replaced. If this time has been exceeded, the device answers automatically with the green LED flashing. This replacement 
of the sensor cartridge can be repeated without limitation, as long as there isn’t a malfunction (fault X4). The evaluation 
unit must not be exchanged. 

After exchange of the sensor, the LED shows the correct operation again (green). 

 

 

Содержание PolyGard 2 MSC2

Страница 1: ...bH W rdinger Str 27 27A 94060 Pocking Germany Specifications subject to change without notice PolyGard is a registered trademark of MSR Electronic GmbH PolyGard 2 MSC2 Multi Sensor Controller Datashee...

Страница 2: ...5 interface with serial protocol For convenient operation the MSC2 is already completely configured and parameterized ex works The intended sites within the ambient conditions as specified in the Tech...

Страница 3: ...Buzzer status LED red yellow green ALARM RELAYS 3 3x Alarm relays POWER SUPPLY 5 90 240 V AC 24 V DC 5 VA 7 90 240 V AC 24 V DC 15 VA HOUSING C Housing type C 130 x 130 x 75 mm ELECTRICAL CONNECTION...

Страница 4: ...W Note Relay 1 2 Status OFF means Alarm off Relay coil energized Status ON means Alarm on Relay coil de energized Note Relay 3 Status OFF means No fault and MSC2 in normal operation Relay coil energiz...

Страница 5: ...n flashing represents a connection option for warning devices to improve visibility If flashing is set this must not be used as a safe output circuit any more A combination of relay mode energized wit...

Страница 6: ...from above the sensor head SC downwards Observe possible constructor s instructions Mounting Controller Wiring The technical requirements and regulations for wiring electrical security fire protection...

Страница 7: ...in the field because it is here where most of the causes for problems in the field bus communication appear Optical Check Right cable type used Correct mounting height according to definition in Moun...

Страница 8: ...100 Vol 2066 R11 R11 ppm 2059 R12 R12 20 2000 ppm 2070 R22 R22 20 300 ppm 2061 R23 R23 20 2000 ppm 2060 R32 R32 ppm 2064 R123 R123 20 300 ppm 2077 R134a R134a 20 300 ppm 2063 R1234yf R1234yf ppm 2065...

Страница 9: ...erial No Date of Production Mainten interval Mainten Password AV Overlay AV Time Power On Time Error Time CFM dupl Note Note down 1900 V time ppm 0 0 0 90 30 30 Analog Output 1 Outp source Oper Signal...

Страница 10: ...1 A2 A3 A4 A1 A2 A3 A4 01 active n CO2 Vol 5 0 5 1 1 1 0 2 0 0 CV 0 0 0 0 1 R1 R2 R3 R4 02 inactive n 03 inactive n Change of values after replugging the jumper closed 01 active n CO2 Vol 5 1 1 8 1 1...

Страница 11: ...utton on the horn the horn acknowledgment must be checked After removal of the reference gas all outputs must automatically return to its initial position Other than the simple functional testing it i...

Страница 12: ...y third parties every customer receives his own internal project key All projects of the customer are delivered with this key The key is also stored in each MSC 06 STL Tool that the respective custome...

Страница 13: ...WH BL24 Tast SPECIFICATIONS SENSOR CARTRIDGE Sensor element Gas type Carbon dioxide CO2 Sensor element Infrared NDIR Measuring range 5 vol Measuring interval 2 sec Accuracy 10 of range Response time t...

Страница 14: ...rol installations and the latest edition of the National Electrical Code ANSI NFPA70 The equipotential bonding required also e g secondary potential to earth or grounding measures must be carried out...

Страница 15: ...MSR Electronic GmbH W rdinger Str 27 27A 94060 Pocking Germany Specifications subject to change without notice PolyGard is a registered trademark of MSR Electronic GmbH...

Отзывы: