7)
To input the C-Cal value,
press
F2
to start the number
entry process.
8)
Use
F2
to change the
number, and
F1
to enter
the number. Add the next
digit by pushing
F2
and
scrolling as required.
Repeat this sequence
until the entire C-Cal
number is entered.
5 times
fixed
blinking
blinking
blinking
blinking
SCROLL
F2
SCROLL
F2
SCROLL
F2
F2
SCROLL
ENTER/SELECT
F1
Error Correction: If you input a wrong value, press
ZERO
to step
back one digit and reenter.
To input a decimal point, push
POWER
while a digit is blinking.
F1
9)
Once the entire C-Cal
value is entered, press
F1
to finalize the number.
The 7300 modifies its
span factor registers to
adjust the calibration to
the value of the C-Cal. If
the C-Cal input was in
the acceptable range, the
LCD will read
PASS
.
Multiple C-Cal span points are
possible, but are only
accessible using MSI’s
optional SCCMP program.
10) When the final C-Cal
number is finished, press
ZERO
to exit and save
the new C-Cal
calibration. The LCD
reads
STORE
to
indicate a successful
calibration.
11) Press
ZERO
again to exit
the CAL menu and start
standard link tension
operation.
blinking
blinking
fixed
fixed
fixed
3DVV
SCROLL
F2
SCROLL
F2
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
blinking
SCROLL
F2
SCROLL
F2
blinking
0
EXIT/SAVE
ZERO
0
EXIT/SAVE
ZERO
6WRUH
SCROLL
F2
blinking
&$/ '
6HWYS
Calibration
33
Figure 5-4 C-Cal Calibration Procedure (continued)