
Calibration
27
Procedure for the Routine Calibration of the MSI-7300
XQ/G
/R$'
lb
Blinks
Capacity
1)
With the power on, initiate
calibration by holding down
the
F2
and
ZERO
keys until
the display reads
CAL
.
The CAL setup menu
appears.
2)
Press
F1
to start the
calibration procedure.
3)
The display reads
UnLd
(unload) indicating you
should remove all weight
from the link.
You can choose to
leave bottom fittings on the link as
long as they are always part of the
load train.
4)
Press
F1
. The 7300 sets the
zero calibration point.
5)
If the zero is in range, the
scale will display
PASS
.
Then
LoAd1
is displayed.
6)
Load the link with a precision
test weight.
For highest accuracy,
a test weight of 10% of capacity or
more is recommended.
7)
Press
F1
. The 7300 flashes
the capacity and calibration
units. If you are loading the
scale with the capacity
weight, skip to step 10.
8)
To enter a calibration
weight other than
capacity, press
F2
. The
displays far left digit will
flash zero indicating that
a number should be
entered.
&$/
blinking
3$66
No weight
on link
Flashes
Zero
Lift Test
Weight
example capacity
Error Correction: If you input a wrong
value, press
ZERO
to step back one digit
and reenter.
ENTER/SELECT
F1
ENTER/SELECT
F1
ENTER/SELECT
F1
SCROLL
F2
Press
Simultaneously
F2
0
ZERO
F1
Figure 5-2 Standard Calibration Procedure
Содержание Dyna-Link 2
Страница 2: ...MSI 7300 Dyna Link 2 Tension Dynamometer Operator s Manual PN 152160 Rev C...
Страница 3: ......
Страница 58: ...Appendix 53 Notes...
Страница 59: ...54 MSI 7300 Operator s Manual Notes...
Страница 60: ......