INSTALLATION
13
2.
INSTALLATION
2.1
Power Supply and Fuses
When the equipment is first received, make sure the voltage selector switch
is correctly positioned to suit your power supply. This switch is located at the
rear of the unit, next to the process timer connection.
Table 2 - Voltage Selector Switch Positions
Voltage
Position
Indication
220/240
Downward
240
120
Upward
120
Two anti-surge fuses are factory fitted in the holders located beneath the mains
inlet connection socket.
Location of the Instrument and Electrical Connections
Place the Soniprep 150 on a suitable table or bench and ensure that at the rear
of the instrument is at least 50mm (2’) away from the wall to allow adequate air
circulation.
See ‘ELECTRICITY SUPPLY SHEET’ (front of manual) for plug/mains cable
details.