DF-8500 Gas Detector
note that the sensors must be protected from exposure to high concentration solvent
vapors for storage or installation and use in the instrument.
Cross Sensitivity:
Cross sensitivity of the oxygen sensor:
There is no cross sensitivity for toxic gas or oxygen sensors under the threshold limit
value (TLV).
However, high concentration (such as over 1%) and strong oxidizing gas
(such as ozone and chlorine) will affect the oxygen equivalent, but most of the othe
r
common gases will not be affected.
DF-8500 cross-sensitivity list for the hydrogen sulfide sensor:
DF-8500 cross-sensitivity list for the carbon monoxide sensor:
Precautions for using DF-8500 ETO, VCM and methanol sensor:
1) Must use polytetrafluoroethylene (Teflon)tube for
calibration,
and the tube is short as
much as possible to prevent adsorbing the gas.
58
Gas
Concentration values
Reading
Acetylene
100
88
Ethylene
100
97
Hydrogen
100
<28
Nitric oxide
48.6
14
Nitrogen dioxide
19.5
<0.5
Chlorine
13.7
<0.5
Ethanol
200
0
Hydrogen sulfide
50
0
Sulfur dioxide
20
0
Ammonia
20
0
Gas
Concentration values
Reading
Carbon monoxide
100
<2
Sulfur dioxide
2
0
Nitrogen dioxide
3
0
Nitric oxide
25
0
Chlorine
0.5
0
Ethylene
100
0
Carbon dioxide
5000
0
Ammonia
50
0
Содержание DF-8500
Страница 3: ...DF 8500 1 1 1 DF 8500 SUPREMA Gasgard XL 9010 9020 8020 1 2 4...
Страница 4: ...DF 8500 1 3 5...
Страница 5: ...DF 8500 1 4 1 2 12 2 2 1 1 DF 8500 6...
Страница 7: ...DF 8500 3 3 1 3 1 1 mm 3 DF 8500 8...
Страница 12: ...DF 8500 4 4 1 4 0mA LED LED CO H2S 1000 1000 mol mol 100 300 100 mol mol 300 mol mol 1 mol mol 1ppm 13...
Страница 20: ...DF 8500 M 14 M 02 M 02 M 14 1 M 03 2 M 14 3 ON 4 4 4 5 M 14 M 03 OFF M 14 ON M 03 6 4 3 0 25 21...
Страница 21: ...DF 8500 4 2 2 M 01 4 2 2 M 02 OFF 4 4 4 0000 1 2 ON 3 4 5 5 1 M 05 22...
Страница 25: ...DF 8500 6 5 DF 8500 1 0 3 DF 8500 25 DF 8500 TLV 1 DF 8500 26 100 2 2 0 3 0 25 0 0 5 0 100 0 5000 0 50 0...
Страница 28: ...DF 8500 9 9 1 29 LED 4 20 mA 4mA 4mA E XX 2mA HI 20 5mA BOOT 0mA 0 3 2mA...
Страница 33: ...DF 8500 SJ T 11364 GB T 26572 GB T 26572 34 Pb Hg Cd Cr VI PBB PBDE DF 8500...