USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
89 /
117
1.Gas
2. Current measured value
3. Setpoint for the measuring channel (in this example N0
2
50ppm).
Reference point calibration
►
Connect the sample gas cylinder to the analyser.
See also Chapter 12.1 Connecting analyser to sample gas cylin-
der, Page 87.
►
Feed gas to the analyser. Wait until the gas concentration no longer
changes. (at least 8 minutes)
►
Write the sample gas value of the sample gas cylinder in the field for
the gas concentration. To write the value in the field, touch the field
and enter the value with the appearing keypad. The value must be
confirmed with the hook.
►
Press the "ADJUST" field. This changes the calibration factor for the
measuring channel.
►
After leaving the calibration window, a button appears in the display.
►
Press "Save?" to permanently save the calibration factors.