USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
13 /
117
1 Sample gas filter (PTFE)
2
Dust filter
3
Auto-zero solenoid valve
4
Sample gas pump
5 Double stage gas cooler
6
Sample flow sensor
7 Paramagnetic oxygen sensor
8
Acrodisc PTFE filter
9 Infrared (NDIR) bank
10 Diff. pressure sensor
3.3.
Analyser
The Analyser consists of a compact and robust metal
housing with shock-absorbing rubber corners. All electrical and
pneumatic connections are located on the both front sides of the
analyser. The analyser is operated exclusively via the touch-sensitive
touch screen.
The analyser is delivered in a water resistent bag.
During measuring operation, the analyser must be placed in the
Water resistant bag. The water-repellent bag is part of the analyser.