16.5.2
Error diagnosis regarding the condensate separator
1. Effect
2. Cause
3. Solution
Dirt and / or humidity inside the
device
No filter effect
Sensor failure
Pump failure
Fine filters are wet and / or dirty.
Check filters more often
Renew them if neces-
sary (white = OK)
Brown-black = renewal
Wrong measuring values
Cover, intermediary unit, plexiglass
tube and locking pieces are not tightly
fixed respectively screwed
Check tightness with
every filter change.