11.5 Operation temperature
If the analyser has been stored at low temperatures, it will require some time to equilibrate to the
ambient temperature before being switched on. If it does not equilibrate, condensation will occur
inside the analyser!
If the temperature is out of its operation range (see chap. 15), you will see the following messages
on the display.
or
Once one of these messages appears you will not be able to use the analyser, it will give an
acoustic signal until it has reached the specified operation temperature b5°C and +45°C.
11.6 Condensate separator
The condensate separator must be checked before and after each measurement!
Please check if the condensate separator must be
emptied a
nd if the star filter is still
white
.
White = good for measurement
dark = replace
11.7 Connectors and leak tightness
•
Check all push on connectors for proper fit.
•
Check all hoses, hose connectors and the condensate separator.
•
Start separate instrument tightness test (see chap. 14.4)