- 6 -
□ TEMP. □HUMIDITY CONTROL OPERATION
1.
Turn on POWER switch (110V or 220V 50/60Hz):
Operation: 1.
Press SET button and SV light will be flashed
2.
Shift (Use up▲ or down▼ to set temp.)
3.
Press SET button when temp. setting complete
4. The machine will begin to run the SV temp.
2.
ALM1 : High temp. warming alarm (AL1)
Operation: 1.
Press SET button after 5 seconds, the AL1 will show on
PV column.
2.
Set high temp. warming degree (Use up▲ or down▼)
3.
Press SET button again after 5 seconds then return to
original setting (PV).
4.
(Using added number on PV to set the high temp. warm
alarm)
Example: Set PV to 37℃, AL1 set 1℃, when the temp. become 38℃, the
Temp. warming alarm will be on and locked the temp. on
38℃ to protect overheat.
(37℃+1℃=38℃
PV +Protecting degree=AL1
3
ALM2: Low temp. warming alarm. (AL2)
Operation:
Example:
1.
Press SET button after 5 seconds, the AL1 will show on
the PV, press SET again and AL2 will be showed.
2.
Set protecting low temp. alarm degree on SV column
(Use up▲ or down▼)
3.
Press SET button again to return original PV.
4.
Using deduct number on PV to set the low temp.
warming alarm)
Set PV to 20℃, AL2 setting -1℃, the warming alarm will
be on when the temp become 19℃ and will be locked the
temp. on 19℃ to avoid the machine be auto-off when the
temp. over low. (20℃-1℃=19℃
PV-Warming degree=
AL2)