- 6 - 

□ TEMP.    □HUMIDITY CONTROL OPERATION

1.

 

Turn on POWER    switch (110V or 220V 50/60Hz):

Operation:  1.

 

Press SET button and SV light will be flashed 

2.

 

Shift (Use up▲ or down▼ to set temp.)

3.

 

Press SET button when temp. setting complete

4. The machine will begin to run the SV temp.

2.

 

ALM1 : High temp. warming alarm (AL1)

Operation: 1.

 

Press SET button after 5 seconds, the AL1 will show on 

PV column. 

2.

 

Set high temp. warming degree (Use up▲ or down▼)

3.

 

Press SET button again after 5 seconds then return to

original setting (PV). 

4.

 

(Using added number on PV to set the high temp. warm

alarm) 

Example:  Set PV to 37℃, AL1 set 1℃, when the temp. become 38℃, the 

Temp. warming alarm will be on and locked the temp. on 

38℃ to protect overheat.   

(37℃+1℃=38℃ 

 PV  +Protecting degree=AL1 

3

 

ALM2: Low temp. warming alarm. (AL2) 

Operation: 

Example: 

1.

 

Press SET button after 5 seconds, the AL1 will show on

the PV, press SET again and AL2 will be showed. 

2.

 

Set protecting low temp. alarm degree on SV column

(Use up▲ or down▼) 

3.

 

Press SET button again to return original PV.

4.

 

Using deduct number on PV to set the low temp.

warming alarm) 

Set PV to 20℃, AL2 setting -1℃, the warming alarm will 

be on when the temp become 19℃ and will be locked the 

temp. on 19℃ to avoid the machine be auto-off when the 
temp. over low. (20℃-1℃=19℃ 

 PV-Warming degree=

AL2) 

Содержание PGI-850H

Страница 1: ...1 Growth Chamber USER S MANUAL PGI 850H RH MRC MRCLAB COM WWW MRCLAB COM...

Страница 2: ...5 2 2Temp Control Operation 6 3 2 1 Temp Setting 7 3 2 2 8 2 3 Timer Control Operation 9 3 3 1 Temp Timer Setting 3 3 2 Lamp Timer Setting 2 4 Over heat low warning lamp 9 3 Trouble shooting for temp...

Страница 3: ...ressor 1 3 Temp Timer Switch Turn on off Temperature Timer 1 4 Lamp Timer Switch Turn on off the lamp Timer 1 5 Day Night Temp Control Display setting and actual temperature for day or night 1 6 Humid...

Страница 4: ...EMPERATURE CONTROL 3 1 Setting temperature First to press the key of the SV key will be in flashing then start to set the temperature The monitor of SV will be in high light then set the required temp...

Страница 5: ...g 3 OUT Output controller signal light 4 AT Micro auto calculate function light 5 SET Change setting parameter 6 SHIFT Change column for setting Temp or RH Hundred digit Ten digit units digit and deci...

Страница 6: ...et the high temp warm alarm Example Set PV to 37 AL1 set 1 when the temp become 38 the Temp warming alarm will be on and locked the temp on 38 to protect overheat 37 1 38 PV Protecting degree AL1 3 AL...

Страница 7: ...column shows symbol or just turn off the power and restart 6 If you are using wider range of temperature please record the parameter value of P I D etc could save your time and key in data directly t...

Страница 8: ...It will heat automatically when the temperature is dropped to the original set point 6 OVER LOW CONTROL Regarding the setting of the OVER LOW CONTROL it generally is set at a temperature lower than 5...

Страница 9: ...ch one notch equal fifteen minute the control axis to outside is change to night temp to inside is change to day temp 2 Lamp Timer Same as Temperature timer Please see below for time switch s manual d...

Страница 10: ...10...

Страница 11: ...11...

Страница 12: ...higher than setting temp b PT platinum damage c thermal meter output message no a wait for heat down or cooling b Replace c ask for repair maintenance Heat light on but temp unable to go up AL1 a hea...

Страница 13: ...urrent input b loose socket c unstable voltage input d fuse bounce e switch indicated light doesn t light f power cord broken or poor contact a check power supply system b plug firmly or replace socke...

Страница 14: ...or RY damage e SENSOR damage a ask for repair maintenance b ask for repair maintenance c re auto tuning check if compressor stop if yes repair d ask for repair maintenance e ask for repair maintenance...

Страница 15: ...ack maintenance e Improve the environment move to the cool place well ventilated environment with ambient temperature 30 f Ask for repair maintenance g Test contactor by hand to see whether it is work...

Страница 16: ...it over 45 angle slope on remove machine you could only use it after 2 4 hours or more to protect the Compressor 6 On settle the machine it must remain more than 30cm to the wall 20cm on two side The...

Отзывы: