6
(5) Unsuitable heat transfer media will negatively affect the
instrumentation and cause damage to the device. Therefore, only use the
manufacturer's specified thermal media and use within the specified
pressure range. The temperature can only be controlled below the boiling
point of the heat conduction medium, and the viscosity must not exceed
50 m2s when the heat conduction medium is operated at a low
temperature. Fill the heat conduction medium at room temperature. Fill
the heat conduction medium slowly. At the same time, ensure that no heat
conduction medium overflows. Protective equipment must be worn when
filling, such as anti-corrosion chemical gloves, goggles, etc.;
(6) The setting parameters to be reconfirmed after filling, which are the
requirements for correct operation and use of the instrument;
(7) Do not tie the heat conduction medium piping;
(8)Check the heat conduction medium tube regularly to prevent material
strain (e.g. fission).
IX. Location
Warning!
1. Ensure that the instrument is vertically upward when moving it;
2 The instrument needs to be in a horizontal plane and fixed;
3.Be put on a surface of a nonflammable object.
4 Keep the environment around the instrument clean and avoid slipping;