11
be replaced regularly to ensure safe use).
(3). To connect with new-fashioned glass reactor: using
stainless steel bellows connection, and the connector adopts
detachable connection.
(4). The device connector is a consumable item and is not
covered by the warranty scope of the whole machine.
2. To fill the heat conduction fluid
a. Before adding the heat conduction fluid to the
equipment, connect the inlet of the heat conduction fluid for
the heating and cooling cycle with the inlet of the heat
conduction fluid at the upper end of the reactor using a
stainless steel corrugated pipe; connect the outlet of the heat
conduction fluid of the high- and low-temperature circulating
machine with the lower end of the reactor using stainless steel
bellows.
b. Turn on the upper exhaust valve above the high-low
temperature cycle machine and open the filling port.
c. Inject the heat conduction oil into the filling port
(at the first time, it can be added to the expansion liquid level
line); turn on the power; press the cycle button (the system
will automatically discharge air); interval 30S, close the
cycle button (during the exhaust process, adding the