6 Advanced Setting
6.1.1 Holding the START button and open the
main switch. After about 5 seconds it will
enter the setting screen.
6.1.2 Select the state by press PROGRAM button.
The state you selected will glitter.
Press the START button to enter the setting.
If you select the S1 and enter the state.
You may change the unit of temperature and
pressure, adjust time and date.
6.2.1 You will select the unit of temperature first.
Press TEMP button to select
or
The unit you select ed will be lighted. Press
the PROGRAM button to the next item.
6.2.2 You may set the pressure unit as the same
method.
6.2.3 Then press PROGRAM button to the next item to adjust the time and
date. After the last word of the date or time is set , then the data is
permitted to be saved. If you want to finish the setting you shall press
START. It will return to the screen of selecting states.
6.1 Enter the setting
6.2 S1 state
℃
.
If use this sterilizer on a place above 2 km, you need to reevaluate the
sterilization result. And you may correct the effect by prolong the
holding time.
The Machine No. And cycle No can not be set by the operator.
6.3 S2 state
6.3.1 You may check the count of sterilization
cycle. It can not be changed by operator.
6.3.2 Set the parameter for high altitude.
If you can't enter the holding time and use
this machine at a high altitude place that is
above 2.0 kilometers or atmospheric
pressure is below 80kPa you need set the
parameter. The scope is 0~2.
6.3.3 Language set:
00 English 01 German
02 Spanish
03 Polish
04 French
05 Magyar
06 Romanian
07 Dutch 08 Lithuanian 09 Latvian
cycle No
machine No
Altitude set
Language
set
1 2
5.5 Test programs (CLASS B)
5.5.1 Press PROGRAM button, select the "B&D TEST".
5.5.1.1 Put the Bowie-Dick pack into the
chamber. Then close the door and
press "START".
5.5.1.2 After finish the cycle you check the
indicator. And evaluate the result.
5.5.2 Select the “HELIX TEST"
5.5.2.1 Put the indicator paper in the capsule.
5.5.2.2 Put the Helix tube into the chamber.
Then close the door and press
"START".
5.5.2.3 After finish the cycle you check the
indicator. And evaluate the result.
5.5.3Select the "VACUUM TEST"
5.5.3.1 Then close the door and press
"START" button.
5.5.3.2 After finish it will show the result.
5.5.3.3 In compliance with EN 13060, the
test requires the air leakage rate
less than or equal 0.13kPa/min
during the 10 minutes.
If the leakage rate is not greater 0.13,
it will show 0 means success.
Or it will show F means failure.
5.5.3.4 If the temperature deference between
the max. temperature and the min.
temperature is above 3, it will show
the value T on the screen and show F.
That means the result of test is void.
You need run the vacuum test again
after the chamber has cooled down.
Caution: The VACUUM TEST must be
carried out with unit cold. If
the Tp is greater 3
,
it will
show failure.
℃
l e a k a g e r a t e
t e m p e r a t u r e
d e f e r e n c e
11
Instructions manual
Instructions manual
F