Recirculating cooler
- 2 -
Chapter II Installation Instruction
The recirculating cooler shall be equipped indoor with good environment. The
principles of choosing the location as following:
Location Principle
1
Firm ground or experiment table in a dust-free and cool
environment with good ventilation
2
Far from droplets or steam
3
Far from the location that influx, viscous flow, leak of fuel gas may
happen
4
Far from high frequency devices such as high frequency welding
machine
5
Far from the location where acidic solutions used frequently
6
Far from the location where special sprayer (sulfide contained)
used frequently
As for convenience and safety of maintenance and installation, space between
the indoor unit and obstacles should be ensured.
Ensure the measurements during indoor installation:
1
)
The distance between the bath top and top obstacles shall be at least 300mm;
2
)
The distance between two bath sides and obstacles shall be at least 500mm;
3
)
The distance between bath back and obstacles shall be at least 500mm;
4
)
The distance between bath front and obstacles shall be at least 1000mm;
The following picture shows the space requirements for installation:
(Measurement unit: mm)