![MRC Rova-M-C3 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/mrc/rova-m-c3/rova-m-c3_users-operating-manual_1845447024.webp)
Recirculating cooler
- 21 -
Item
Parameter Meaning
Option
or
Range
Initial
Data
11
-
E_CH
alarm
End
_END
Parameter set end signal
_ _ _ _
_ _ _
_
Appendix II Warning Information
Item
Alarm lamp
and buzzer
Auxiliary
display
Main
display
Solution
High temperature
alarm 1
2S flash
E_H1
Current
temperature
No,
quit
when
the
temperature lower than
E_H1, back to original work
mode
High temperature
alarm 2
1S flash
E_H2
Current
temperature
No,
quit
when
the
temperature lower than
E_H2, back to original work
mode
Low temperature
alarm 1
2S flash
E_L1
Current
temperature
No,
quit
when
the
temperature higher than
E_L1, back to original work
mode
Low temperature
alarm 2
1S flash
E_L2
Current
temperature
No,
quit
when
the
temperature higher than
E_L2, back to original work
mode
Water level alarm
2S flash
E_LW
Current
temperature
Stop heating/cooling, quit at
normal water level and back
to original work mode
Water flow alarm
1S flash
E_LF
Current
temperature
Stop heating/cooling, quit at
normal water level and back
to original work mode
Sensor 1 break
1S flash
E_S1
999
Stop heating/cooling, quit at