Recirculating cooler
- 11 -
adjusting temperature. Ordinarily it is about 15 to 20 minutes.
3.3.6 Switching off
1. Press Start/Stop key
to stop the compressor, the system enters
into standby state, then press Pump key
to turn off then pump.
2. Switch off the power supply of the control panel;
3. Please turn off the electric leakage protection and power supply switch,
empty and dry the water tank and pump for long time outage;
4. After long time usage, the pump should be opened to check and change
the worn-out parts when water flow and pressure decrease obviously;
5. Please ensure no water in the unit (especially in the pump and plate heat
exchanger) when the environment temperature is under 0
℃
. Residual
water will freeze and do damage to parts such as the pump, heat
exchanger and pipes. It is suggestive to use antifreeze as circulating
medium for long time operation and outage in low temperature
environment.
3.4 Advanced application mode
Long press Menu key
over 3 seconds into advanced application mode.
The main LED digital display shows four flashing slashes prompting for code,
press Add key
and Reduce key
to input data, press Left-shift
key
to move the input location, press Enter key
to enter into
parameter set mode after finished inputting code. Different parameter set
interface can be loaded with inputting different code.
Take setting operation model base as an example in the following text (The
other parameter database setting method is similar, please refer to Appendix