background image

  

11 

Display Screen   

S

witch on the power, the screen will first show

 

the version and model name. 

After the welcome page you will be able to select the operation mode: 

Constant

 or 

Program mode 

(Screen shown below)   

Constant Setup 

     

Program Setup 

 

 

Constant Setup 

 

This section gives an instruction for constant mode operation. 

Constant 

Voltage / Current / Power Mode 

allow you to specify a voltage limit or a 

current limit to be used continuously during the entire duration of 

electrophoresis.  Review the guidelines provided in this manual before 

starting electrophoresis using 

PWR-3030

 Power Supply. We recommend reading 

the guidelines provided in this manual for best results before starting an 

operation. 

 

1. Place the 

PWR-3030

 Power Supply on a sturdy and level surface in a safe and 

dry place, away from laboratory traffic.   

 

2. Ensure that the AC power switch is OFF, and then plug the three-pronged 

power cord into a grounded three-prong AC outlet of the appropriate voltage 

(110V to 240V as indicated on the rating sticker near the AC cord on the 

back of the unit). 

 

3. Connect the DC output jacks from the electrophoresis unit; insert the red 

lead (+) into the red output jack, and the black lead (-) into the black output 

jack. 

 

Содержание PWR-3030

Страница 1: ...PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC VER 01 2 12 Operation Manual for Power Supply PWR 3030...

Страница 2: ...MRC is liable for all missing or damaged parts accessories within 7 days after customer receives this instrument package Please contact MRC immediately regarding this issue If no response is received...

Страница 3: ...scription 7 Introduction 8 Overview 8 Controls and Features 9 Installation Instructions 10 Operation Instructions 10 Display Screen 11 Constant Setup 11 Programming Setup 14 Limiting Parameter Setting...

Страница 4: ...e following points carefully before operating this equipment 1 Read and follow the manual instructions carefully 2 Do not alter the equipment Failure to follow these directions could result in persona...

Страница 5: ...ollution degree 2 7 Mains supply voltage fluctuations up to 10 of the normal voltage Avoiding Electrical Shock Follow the guidelines below to ensure safe operation of the unit PWR 3030 power supply ha...

Страница 6: ...et the unit equilibrate to room temperature for a minimum of 2 hours before using it Equipment Operation Follow the guidelines below to ensure safe operation of the unit 1 Check the displayed figures...

Страница 7: ...nstant mode 1 9999 min with alarm continuous Programmable mode 1 999 min with alarm continuous Safety Device No Load detection Leakage detection Sudden load change detection Over temperature protectio...

Страница 8: ...multiple electrophoresis units simultaneously MRC PWR 3030 power supply provides Constant Voltage or Constant Current or Constant Power to instruments used in electrophoresis 4 pairs of terminals and...

Страница 9: ......

Страница 10: ......

Страница 11: ...constant current modes adjust the other parameter to the maximum value For example to operate under constant voltage adjust current to max before running using constant voltage and vice versa The PWR...

Страница 12: ...provided in this manual before starting electrophoresis using PWR 3030 Power Supply We recommend reading the guidelines provided in this manual for best results before starting an operation 1 Place th...

Страница 13: ...wer W or time Min Press the key to set the specified parameter 7 Use or to set the appropriate value and then press Move to the next parameter until all the parameters are set in the same operation me...

Страница 14: ...wer OFF by the switch on the rear 12 To change limits of electrophoresis in progress If you need to make changes to the current running limits you must stop electrophoresis by pressing the Key Press K...

Страница 15: ...ting an operation 1 Place the PWR 3030 Power Supply on a sturdy and level surface in a safe and dry place away from laboratory traffic 2 Ensure that the AC power switch is OFF and then plug the three...

Страница 16: ...value and then press and move to the next parameter until all the parameters are set in the same operation method 9 Use Key to move down to for setting Step 4 6 Press Key back to Step 1 3 screen 10 Pr...

Страница 17: ...sis in progress If you need to make changes to the current running limits you must stop electrophoresis by pressing the Key Press Key to enter the setting screen Enter the changes and then press Key o...

Страница 18: ...ng 1 Use Key Key Key Key and Key to set Maximum Current 3A and Maximum Power 300W and on the real time screen Volt will be shaded for instance Current Limiting 1 Use Key Key Key Key and Key to set Max...

Страница 19: ...tops with alarm The screen displays Electrophoresis leads are not connected to the power supply or to the electrophoresis unit s or there is a broken circuit in the electrophoresis cell Check the conn...

Страница 20: ...nnect the power cord from the outlet and contact Technical Service Operation stops with alarm Display shows Ground leak detected during run Check the electrophoresis system for improper grounding Rest...

Страница 21: ...wer Supply and detach the power cord from the rear of PWR 3030 Power Supply 2 Open the fuse compartment located inside the Power Entry Module by inserting a small flathead screwdriver into the slot be...

Страница 22: ...intenance or recalibration under normal use The housing may be cleaned with a dry cloth If the unit must be returned for repair contact our SERVICE DEPARTMENT or your local distributor for shipping in...

Страница 23: ...rranty excludes damages resulting from shipping misuse carelessness or neglect MRC s liability under the warranty is limited to the receipt of reasonable proof by the customer that the defect is embra...

Отзывы: