Operation
Manual
Chapter 2
Features
2
Chapter 2 Features
2.1.
Normal Operating Conditions
Ambient temperature range: 10
℃
~35
℃
, Relative humidity range: 20%~75%;
The atmospheric pressure: 860hPa~1060hPa;
Power supply: Input AC 100~240V, 50Hz~60Hz; Output DC15V, 8A;
Ambient: Dry, clean, flat and firm horizontal surface, Away from direct sunlight,
mechanical vibration and strong magnetic field interference.
2.2.
Basic Parameters and Performance
Technical
parameters
Product model
PCR-IF-8C
Capacity
0.2mL–8wells
(Strip of 8 tubes)
Reaction system
10-150μL
(Recommended 20-30μL)
Power supply/Power Max
AC100~240V,50~60Hz/100VA
Dimensions(W*D*H) Max
145*305*100mm
Weight Max
2.5kg
Temperature
control
Heating lid function
Yes, temperature is adjustable
Temperature range(
℃
)
15
℃
~99
℃
Maximum heating rate(
℃
)
≥2
℃
/s
Maximum cooling rate(
℃
)
≥2
℃
/s
Temperature uniformity(
℃
)
±0.15
℃
@39
℃
Temperature accuracy/
Control accuracy(
℃
)
±0.2
℃
/ 0.1
℃
Optical source
Applicable dye
FAM/SYBR Green
Ⅰ
, SYTO-16, etc.
Excitation wavelength
Ex470nm/LED
Detection wavelength
Em535nm/PD
Function
Display
7-inch touch screen
Basic function
Program and sample parameter editing,
fluorescence acquisition, hole negative
positive and peak parameters setting,
normalization function, data management,
data exporting, amplification curve, etc.
Содержание PCR-IF-8C
Страница 2: ...I...