LCEN-401P

14 

·

Samples are imbalance

symmetrically 

A

lar

m

 code

 appear

ed on t

h

tim

di

spl
ay s

cr

een

 

E-02

Door fault 

·

The door opened in running.

·

The 

button  is  pressed 

while the door opening. 

·

Close the door immediately.

·

Close the door

and then start 

to operate. 

E-06

Set wrong speed 

·

The  setting  speed  exceed  the

allowable range. 

·

Modify the speed value.

E-10

86

·

Read the service manual

·

Contact with service center

Table 11-1 Possible problems and solutions 

Alarm codes E-1

E-9 are related to incorrect operation/programming. You can continue running the

centrifuge after implementing corrective procedures.

11.2

 

How to open the door 

1) In the case of

 

power on

1

Turn on the power switch, release the door automatically.

2

The door will be released automatically once the operation is finished.

3

It is available to release the door by press

button once the rotor stops. 

2) In the case of power outage

The door cannot be opened automatically if there is a power outage. It is available to be opened manually 

as follows. 

1

Ensure if the rotor has stopped rotating. 

Listen carefully to ensure no rotating sound can be heard.

2

Insert a screw driver into a hole to open door.

Holes are located on the left and right sides of the unit.

Insert a screw driver into the two holes and push forward to release the door.

12. Instructions for the rotor and tubes

CAUTION 

The door just can be opened while the power is on and rotor stops rotating.

Содержание LCEN-401P

Страница 1: ...LCEN 401P PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 2 22 PRP PRF CGF CENTRIFUGE...

Страница 2: ...tubes balanced 8 8 4 Inspect the rotor 8 8 5 Symmetrically load centrifuge tubes into rotor 8 9 Operation 8 9 1 Normal operation 8 9 2 RCF operation 11 9 3 Reset 11 10 Maintenance 11 10 1 Cleaning 11...

Страница 3: ...n personal injury or possible death CAUTION Possible damage to centrifuge Caution notes indicate any condition or practice which if not strictly observed or remedied could result in damage or destruct...

Страница 4: ...entrifuge except by our service center are strictly prohibited CAUTION This centrifuge must be located on a firm and level table Make sure the centrifuge is horizontal before running Do not move or re...

Страница 5: ...ncrement 10 g Maximum capacity 10ml 12 15ml 8 Timer 30seconds 99minutes Noise 56dB A Driving Motor Brushless DC motor Safety devices Door interlock Over speed detector Error code runtime display Power...

Страница 6: ...ital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This...

Страница 7: ...firmly Avoid installing on a slippery surface or surface prone to vibration 2 Ideal ambient temperature is 20 5 avoid placing the centrifuge in direct sunlight if temperature exceeds 30 3 Keep clear...

Страница 8: ...Figure 6 1 Front view of the centrifuge Figure 6 2 Rear view of the centrifuge 7 Operation panel Door lock hook Rotor Operation panel Door release hole LCD Display Power inlet Power switch Air vents V...

Страница 9: ...on 5 Parameter knob Clockwise rotate to increase program values Rotate anti clockwise to decrease parameter values Press the button 5s change the preloaded program parameters Figure 7 2 The main inter...

Страница 10: ...rifuge tubes into rotor CAUTION Make sure the rotor lid is securely fixed on the rotor as well as the rotor and shaft are tightened Otherwise the rotor may be moved off while rotating and cause damage...

Страница 11: ...the speed setting value is reached the screen displays the remaining run time 2 View and modify the operation programs Press parameter knob for 5 seconds When flashing rotate parameter knob to modify...

Страница 12: ...ram operation Press the button to select P program There are 9 programs P1 P9 available and 6 steps for each program 1 P1 P9 programs 1 Press the button to select P1 program When P1 icon flash press p...

Страница 13: ...tep button for 5 seconds and the speed symbol will flash into RCF value input status press again for 5 seconds to return RPM value input mode Rotate program knob to input a RCF value RCF increment is1...

Страница 14: ...ase 4 Door Clean and sterilize the door using the same method as the section 1 above 5 Rotor To prevent corrosion remove the rotor from rotor chamber If not in use for a long term then detach the roto...

Страница 15: ...elow Symptom Causes Solutions Nothing appears on the screen when the POWER is turned on Building power circuit breaker trips Remove the trouble and turn on the POWER Abnormal vibration Rotor do not ma...

Страница 16: ...open the door 1 In the case of power on 1 Turn on the power switch release the door automatically 2 The door will be released automatically once the operation is finished 3 It is available to release...

Страница 17: ...ructure 2 Available rotors and adapters Rotor type Tubes Tubes Per Rotor Dimension L mm Adapters Maximum speed rpm Radius cm Maximum RCF g A12 10P 15ml con 8 17 120 4500 11 2490 1 5 5ml vacu 12 13 82...

Страница 18: ...cidic higher than pH5 O O O Alkaline higher than pH9 O X O Alkaline pH9 or lower O O O Neutral pH7 O O O Warm water up to 70 O O O Ultrasonic cleaning Neutral detergent pH7 O O O Sterilization Autocla...

Страница 19: ...perature applied and so on When the plastic tubes are used for ordinary aqueous samples pH between 5 0 and 9 0 their life expectancies are defined as follows Be operated at the maximum speed High qual...

Страница 20: ...blem 15 2 Warranty of the rotor Plastic rotor lifetime is usually 3 years and we offer 3 years warranty from the date of delivery upon manufacturer Please pay attention do not use the rotor once it ha...

Отзывы: