LCEN-401P
12
direct exposure.
⚫
If the centrifuge needs cleaning, clean it with a cloth or sponge moistened with a neutral detergent
solution.
⚫
Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
(
2
)
Rotor chamber
⚫
If the rotor needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution.
Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
(
3
)
Drive shaft
⚫
We recommend regular maintenance for drive shaft. You can wipe the drive shaft with soft cloth, and
then apply a thin coat of silicon grease.
(
4
)
Door
⚫
Clean and sterilize the door using the same method as the section
(
1
)
above.
(
5
)
Rotor
⚫
To prevent corrosion, remove the rotor from rotor chamber. If not in use for a long term, then detach
the rotor lid and turn upside down to dry the tube holes and keep clean.
⚫
For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor
when it is completely dry.
⚫
The rotor should be checked every 3 months to ensure the tube and rotor holes keep are clean and
apply a thin coat of silicon grease.
10.2 Rotor Installation
(
1
)
Installation
①
②
CAUTION
⚫
Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber, otherwise
fluids may leak into the drive units and cause corrosion or deterioration to the bearings.