2-2 Electrical Specifications (23. 0 ± 5
℃
)
Current source
Current source
Range
Display
Accuracy
Resolution
0.05 - 24.00 mA
0.01 mA
±
( 0.25 % FS )
* Output 0-20 mA current for loads up to 500 ohms.
Output > 20 mA current for loads up to 400 ohms.
* FS : full scale
Voltage source
Voltage source
Range
Display
Accuracy
Resolution
-199.9 mV to +199.9 mV
0.1 mV
±( 0.25 % FS +0.2mV)
* Output measured load inpedence should > 1 K ohms.
Type k source
Type k source
Range
Display
Accuracy
Resolution
-199.9 to 1370.0
℃
0.1 degree
> -100
℃
±( 0.1 % FS + 1
℃
)
< -100
℃
±( 0.1 % FS + 2
℃
)
-327.0 to 2498.0
℉
0.1 degree
> -148
℉
±( 0.1 % FS + 2
℉
)
< -148
℉
±( 0.1 % FS + 4
℉
)
* intput inpedence : 10 ^12 ohms.
Remark :
The above specification are tested under the environment
RF Field Strength less than 3 V/M & frequency less than the
30 MHz only.
3