background image

UNI MP100 User’s Guide 

 

18 

 

Remove the monitor cover and battery only in area known as non-hazardous. 

 

Use  only  mPower’s  lithium  battery  part  number 

M500-0001-000  [1.17.02.0002]

  (3.6V, 

2700mAH, AA size) or part No. ER14505 cell manufactured by EVE Energy Co., LTD. 

  AVERTISSEMENT ! 

 

N'utilisez jamais le moniteur lorsque le couvercle est enlevé. 

 

Retirer  le  couvercle  du  moniteur  et  la  batterie  uniquement  dans  une  zone  connue  comme  non 
dangereuse. 

 

Utilisez uniquement la batterie au lithium de mPower, pièce No. 

M500-0001-000 [1.17.02.0002] 

(3.6V, 2700mAH, taille AA) ou celle de EVE Énergie Cie., Lt

é

, pièce No. ER14505. 

7.2

 

Sensor Filter Replacement 

A “peel-and-stick” filter should be used on the MP100 in order to keep debris from fouling the 

sensor. Replace the filter whenever it appears dirty, is clogged with particles, has contacted liquid, 

or  when  sensor  response  becomes  weak  and/or  slow.    Use  external  clip-on  filters  when 

operating in dusty environments for easier filter exchange.     

 

1)

 

Turn  off  the  MP100  and  remove  the  top  cover  as  described  above  for 

battery replacement. 

2)

 

Peel  a  filter  from  its  sheet,  center  it  over  the  sensor,  and  gently  press 

down.   

3)

 

Reconnect the buzzer and reinstall the top cover as described above for 

battery replacement. Be careful to not overtighten the screws. 

7.3

 

Sensor Replacement 

MP100  models  are  designed  for  easy  sensor  replacement.    CO  and  H

2

S  sensors  have  typical 

operating lives  of 5 years,  while others  are 1 to  2 years,  as  per warranty  (See  Specifications  in 

Section 7.8). 

1)

 

Turn off the MP100 and remove the top cover as described above for battery replacement. 

2)

 

Replace the old sensor with a new one. Make sure the pins are not bent or corroded. Align the 

pins  to  the  corresponding  holes  and  push  the  sensor  straight  in.  The  sensor  should  fit  flush 

against the printed circuit board. 

3)

 

Check the instrument filter and, if needed, replace as described in the previous section. 

4)

 

Reconnect the buzzer and reinstall the top cover as described above for battery replacement. 

Be careful to not overtighten the screws. 

 

 

 

 

CAUTION

!

     

External Filter Clip

 

Содержание UNI MP100

Страница 1: ...UNI Multi Gas Detectors MP100 User s Guide Rev 1 11 October 2019...

Страница 2: ...nfigurations 9 4 6 1 Alarm Limits 9 4 6 2 Span Value 10 4 6 3 Bump Cal Intervals 10 4 6 4 Gas Concentration Unit 10 4 6 5 Vibrator Enable Disable 10 4 6 6 Power on Zero Enable Disable 10 4 6 7 Fast Po...

Страница 3: ...omme non dangereuse Utilisez uniquement la batterie au lithium de mPower pi ce No M500 0001 000 1 17 02 0002 3 6V 2700mAH taille AA ou celle de EVE nergie Cie Lt pi ce No ER14505 Cet instrument n a pa...

Страница 4: ...nd these can be displayed on demand High Low STEL and TWA alarm thresholds are configurable The shell is made of high strength durable material The two key operation is simple to use Sensor and batter...

Страница 5: ...rm action 3 Unit status indicator OK and to confirm entry 4 Gas unit includes x10 6 ppm mg m3 mol mol 5 Battery charge status 6 HIGH LOW STEL TWA alarm indicator when flashing 7 Span calibration in pr...

Страница 6: ...essed to acknowledge otherwise the instrument will turn itself off automatically after 15s Lastly the following values will be shown accordingly High alarm threshold Low alarm threshold STEL short ter...

Страница 7: ...ar Peak screen and Press the Left Key again to acknowledge and clear the Peak value 4 3 5 Minimum Oxygen Sensor Only The Minimum screen is used for the oxygen sensor only and shows the lowest value si...

Страница 8: ...o exit Config Mode press the Right Key until is displayed and acknowledge with the Left Key to return to Normal Mode 4 5 Sensor Calibration and Bump Test Before the unit can monitor gas correctly it n...

Страница 9: ...t 0 3 LPM but no more than 0 6 LPM Use as short tubing connections as possible Span Calibration Procedure 1 Connect the Calibration Adapter to the span gas cylinder s regulator and snap it into place...

Страница 10: ...s 4 6 1 Alarm Limits All the preset alarm limits HIGH LOW STEL TWA can be changed From these menus press the Left Key to change the corresponding alarm limit using the same process as for entering a p...

Страница 11: ...ntly selected unit blinking Unit options include x10 6 ppm mg m3 and mol mol for toxic gas sensors and for oxygen Use the Right Key to scroll through the unit list and select and the Left Key to confi...

Страница 12: ...e version number and then goes directly to concentration readings The fast power on menu alternates between and Press the Left Key to change the fast startup enable disable status Enable or disable Fa...

Страница 13: ...ironments 2 Turn on the instrument and insert it face down into the Docking Box 3 Start mPower Suite on the PC and click the Search button on the bottom panel 4 Find the instrument in the left bar Dev...

Страница 14: ...n be used for calibrations it must be set up for the desired gas type and span concentration 1 Connect the USB cable to both the Docking Box and the PC WARNING Connect only in non hazardous environmen...

Страница 15: ...date entered 7 Hibernate Timeout is the number of seconds of inaction before the Docking Box turns itself off automatically Press the Cal button to turn back on 8 To download Cal Bump test reports cli...

Страница 16: ...ounds connect the air inlet to a fresh air source 3 If desired connect tubing to the gas outlet to exhaust away from the operator breathing area Docking Box Components Calibration Gas Connections 4 Pl...

Страница 17: ...ne Cal bump testing Green Beep Once Cal bump test pass Orange None Sensor type mismatch Red 3 beeps per sec Cal bump test fail Status LED 4 Green None Power On Green blinking None Low battery Orange N...

Страница 18: ...gers every second The battery needs to be replaced as follows 1 Turn off the MP100 and place it face down on a soft surface 2 Use a T10 Torx screwdriver to loosen each of the four screws by turning th...

Страница 19: ...onments for easier filter exchange 1 Turn off the MP100 and remove the top cover as described above for battery replacement 2 Peel a filter from its sheet center it over the sensor and gently press do...

Страница 20: ...ilter Use external filter clip in dusty environments Weak sensor Have Service Technician check raw counts and replace sensor as needed Calibration Adapter is attached Remove Calibration Adapter Readin...

Страница 21: ...1 Vibration per second HIGH 2 flashes per second Low alarm Buzzer 2 beeps per second LED 2 flashes per second 1 Vibration per second LOW 2 flashes per second STEL alarm Buzzer 1 beeps per second LED 1...

Страница 22: ...minute LED 1 flash per minute 1 Vibration per minute Battery Low alarm Buzzer 1 beep per second LED 1 flash per second bAT LoW 1 flash per second Battery Empty alarm Buzzer 1 beep per minute LED 1 fl...

Страница 23: ...0 250 1 25 25 50 25 25 Orange 30 NO2 0 20 0 1 5 1 10 1 1 Orange 30 PH3 0 20 0 01 5 1 2 1 0 3 Orange 60 SO2 0 20 0 1 5 2 10 5 2 Orange 15 ETO 0 100 0 1 10 2 5 2 1 Orange 120 0 200 0 1 10 2 5 2 1 Orange...

Страница 24: ...TWA alarms adjustable Over range alarm Low battery alarm Alarm Signal 95 dB 30 cm Bright red LEDs Built in vibrator Calibration 2 point calibration zero and span power on zero user selectable Event Lo...

Страница 25: ...UNI MP100 User s Guide 24 Technical Support and mPower Contacts mPower Electronics Inc 3046 Scott Blvd Santa Clara CA 95054 Phone 408 320 1266 Fax 669 342 7077 info mpowerinc com www mpowerinc com...

Отзывы: