UNI MP100 User’s Guide
9
4.5.3
Bump Test
A Bump Test is a quick check to ensure that the sensor and alarms are working properly. It is
done with the same gas as is used for span calibration. Enter the
menu, start the gas flow,
then press the Left Key to start bump count-down (typically 45 seconds, but varies with sensor).
After count-down, the bump test result
or
is displayed.
To abort the bump test during count-down, press the Right Key and
is displayed.
Although a Bump Test is a recorded event in the datalog, the user can always do an unrecorded
bump check such as by breathing into an oxygen monitor just to verify that the sensor and alarms
are functioning.
4.6
Setting Instrument Configurations
4.6.1
Alarm Limits
All the preset alarm limits, HIGH, LOW, STEL & TWA can be changed. From these menus:
,
,
,
, press the Left Key to
change the corresponding alarm limit, using the same process as for entering a password (Section
5.1.1):
The current setting value is displayed, with the first digit flashing:
Use the Left Key to increase the current digit, cycling from 0 to 9:
Use the Right Key to move the cursor to the next digit:
After all digits are entered, use the Right Key to move to the “OK” symbol,
and press the Left Key to save the entry. The unit will display SAVE for a
few seconds while storing the value but it is not necessary to press OK to
initiate saving.
NOTE:
The
MP100 will show an error message “Err” if:
•
The Low alarm is attempted to be set higher than the high alarm setting.
•
The High alarm is attempted to be set lower than the low alarm setting.
•
The entered value is outside the measuring range.