UNI MP100 User’s Guide 

 

10 

4.6.2

 

Span Value 

The span gas concentration can be changed from the Cal SET menu using 
the same process as for setting alarm limits. 

 

 

NOTE:

 The

 

MP100 will show an error message “Err” if:   

• 

The  Span  setting  is  less  than  5%  of  the  measuring  range  or  greater  than  the  measuring 
range. 

• 

For the Oxygen sensor, the span setting is greater than 19.0%. 

 

4.6.3

 

Bump/Cal Intervals 

In the Bump and Cal Interval menus, the LCD alternates between: 

  and 

, or 

  and 

. Press the Left Key to enter the menu and change the interval using the same 

process as for setting alarm limits.    Note that a value of 0 means Bump or Cal notifications are 

turned off. 

NOTE:

 The

 

MP100 will show “Err” if the interval is out of the valid range: 0-180 day(s). 

4.6.4

 

Gas Concentration Unit 

The gas concentration unit menu alternates between              and              . Press the Left Key to 

enter the gas unit sub-menu, showing the currently selected unit blinking. Unit options include 

x10

-6

, ppm, mg/m

3

 and µmol/mol for toxic gas sensors, and % for oxygen. Use the Right Key to 

scroll through the unit list and select, and the Left Key to confirm and exit. 
 

4.6.5

 

Vibrator Enable/Disable 

The vibrator consumes a lot of power and can be disabled to extend the battery life. The Vibrator 

menu  alternates  between 

and 

.  Press  the  Left  Key  to  change  the  vibrator 

enable/disable  status.  The  current  vibrator  status  is  displayed,  alternating  between 

  and 

  if enabled, or between 

  and 

, if disabled. Use the Right Key

 

to change the 

status, and use the Left Key to confirm and exit. 
 

4.6.6

 

Power-on Zero Enable/Disable 

The sensor baseline may shift due to changes in environment conditions, such as temperature or 

humidity, and require a zero calibration. The MP100 can prompt the user to zero calibrate every 

time the unit is powered on, and this feature can be enabled/disabled.   

The Power-on Zero menu alternates between 

  and 

. Press the Left Key to change 

the  power-on  zero  enable/disable  status.  The  current  status  is  displayed,  alternating  between 

Содержание UNI MP100

Страница 1: ...UNI Multi Gas Detectors MP100 User s Guide Rev 1 11 October 2019...

Страница 2: ...nfigurations 9 4 6 1 Alarm Limits 9 4 6 2 Span Value 10 4 6 3 Bump Cal Intervals 10 4 6 4 Gas Concentration Unit 10 4 6 5 Vibrator Enable Disable 10 4 6 6 Power on Zero Enable Disable 10 4 6 7 Fast Po...

Страница 3: ...omme non dangereuse Utilisez uniquement la batterie au lithium de mPower pi ce No M500 0001 000 1 17 02 0002 3 6V 2700mAH taille AA ou celle de EVE nergie Cie Lt pi ce No ER14505 Cet instrument n a pa...

Страница 4: ...nd these can be displayed on demand High Low STEL and TWA alarm thresholds are configurable The shell is made of high strength durable material The two key operation is simple to use Sensor and batter...

Страница 5: ...rm action 3 Unit status indicator OK and to confirm entry 4 Gas unit includes x10 6 ppm mg m3 mol mol 5 Battery charge status 6 HIGH LOW STEL TWA alarm indicator when flashing 7 Span calibration in pr...

Страница 6: ...essed to acknowledge otherwise the instrument will turn itself off automatically after 15s Lastly the following values will be shown accordingly High alarm threshold Low alarm threshold STEL short ter...

Страница 7: ...ar Peak screen and Press the Left Key again to acknowledge and clear the Peak value 4 3 5 Minimum Oxygen Sensor Only The Minimum screen is used for the oxygen sensor only and shows the lowest value si...

Страница 8: ...o exit Config Mode press the Right Key until is displayed and acknowledge with the Left Key to return to Normal Mode 4 5 Sensor Calibration and Bump Test Before the unit can monitor gas correctly it n...

Страница 9: ...t 0 3 LPM but no more than 0 6 LPM Use as short tubing connections as possible Span Calibration Procedure 1 Connect the Calibration Adapter to the span gas cylinder s regulator and snap it into place...

Страница 10: ...s 4 6 1 Alarm Limits All the preset alarm limits HIGH LOW STEL TWA can be changed From these menus press the Left Key to change the corresponding alarm limit using the same process as for entering a p...

Страница 11: ...ntly selected unit blinking Unit options include x10 6 ppm mg m3 and mol mol for toxic gas sensors and for oxygen Use the Right Key to scroll through the unit list and select and the Left Key to confi...

Страница 12: ...e version number and then goes directly to concentration readings The fast power on menu alternates between and Press the Left Key to change the fast startup enable disable status Enable or disable Fa...

Страница 13: ...ironments 2 Turn on the instrument and insert it face down into the Docking Box 3 Start mPower Suite on the PC and click the Search button on the bottom panel 4 Find the instrument in the left bar Dev...

Страница 14: ...n be used for calibrations it must be set up for the desired gas type and span concentration 1 Connect the USB cable to both the Docking Box and the PC WARNING Connect only in non hazardous environmen...

Страница 15: ...date entered 7 Hibernate Timeout is the number of seconds of inaction before the Docking Box turns itself off automatically Press the Cal button to turn back on 8 To download Cal Bump test reports cli...

Страница 16: ...ounds connect the air inlet to a fresh air source 3 If desired connect tubing to the gas outlet to exhaust away from the operator breathing area Docking Box Components Calibration Gas Connections 4 Pl...

Страница 17: ...ne Cal bump testing Green Beep Once Cal bump test pass Orange None Sensor type mismatch Red 3 beeps per sec Cal bump test fail Status LED 4 Green None Power On Green blinking None Low battery Orange N...

Страница 18: ...gers every second The battery needs to be replaced as follows 1 Turn off the MP100 and place it face down on a soft surface 2 Use a T10 Torx screwdriver to loosen each of the four screws by turning th...

Страница 19: ...onments for easier filter exchange 1 Turn off the MP100 and remove the top cover as described above for battery replacement 2 Peel a filter from its sheet center it over the sensor and gently press do...

Страница 20: ...ilter Use external filter clip in dusty environments Weak sensor Have Service Technician check raw counts and replace sensor as needed Calibration Adapter is attached Remove Calibration Adapter Readin...

Страница 21: ...1 Vibration per second HIGH 2 flashes per second Low alarm Buzzer 2 beeps per second LED 2 flashes per second 1 Vibration per second LOW 2 flashes per second STEL alarm Buzzer 1 beeps per second LED 1...

Страница 22: ...minute LED 1 flash per minute 1 Vibration per minute Battery Low alarm Buzzer 1 beep per second LED 1 flash per second bAT LoW 1 flash per second Battery Empty alarm Buzzer 1 beep per minute LED 1 fl...

Страница 23: ...0 250 1 25 25 50 25 25 Orange 30 NO2 0 20 0 1 5 1 10 1 1 Orange 30 PH3 0 20 0 01 5 1 2 1 0 3 Orange 60 SO2 0 20 0 1 5 2 10 5 2 Orange 15 ETO 0 100 0 1 10 2 5 2 1 Orange 120 0 200 0 1 10 2 5 2 1 Orange...

Страница 24: ...TWA alarms adjustable Over range alarm Low battery alarm Alarm Signal 95 dB 30 cm Bright red LEDs Built in vibrator Calibration 2 point calibration zero and span power on zero user selectable Event Lo...

Страница 25: ...UNI MP100 User s Guide 24 Technical Support and mPower Contacts mPower Electronics Inc 3046 Scott Blvd Santa Clara CA 95054 Phone 408 320 1266 Fax 669 342 7077 info mpowerinc com www mpowerinc com...

Отзывы: