The reading displayed during the STEL measurement is not the instantaneous reading, but rather the 
accumulated STEL calculation. This is done so that the user can abort work or don protective 
equipment to avoid further exposure if STEL has been reached before 15 minutes have elapsed. 

 

STEL    =    Average Concentration    x   

 
However, the unit still alarms if the instantaneous concentration exceeds the programmed Low or 
High alarm limits. STEL measurement can be aborted by pressing the Right key, and the Right key 
again when the unit beeps, and the Right Key a third time to return to benzene tube sampling. 

CAUTION! 

After a STEL measurement, the pump stops to avoid corrosive liquids from the tube being drawn 
into the monitor. The monitor beeps once per second to notify the user that STEL is complete. Press 
the Right Key to stop the alarm and a warning appears to remove the tube. Use the Left key to 
escape back to Continuous Benzene Screening. 

 

Benzene Tube Replacement 

In general it is good practice to replace the benzene filtering tube after each measurement. When used 
in a high VOC background, and especially after a STEL measurement, it is likely that the capacity of 
the tube to remove interfering chemicals is used up. If the tube has turned from a bright orange to a 
pale orange color, or a greenish-brown stain is seen for more than 2/3 of the tube length, the capacity 
of the tube is exhausted and a new tube should be used for the next measurement. The image below 
shows examples of new versus used tubes. The green-brown-blue colors are formed by reactions 
with VOCs while the pale orange color is caused by absorption of humidity, which in turn reduces 
the ability of the tube to absorb VOCs. Similarly, do not crack open tubes until immediately before 
use, because ambient humidity diffusing in will reduce the capacity for VOCs. Finally, if a tube has 
registered a benzene reading but does not show a discoloration due to other VOCs or humidity, it 
should still be replaced because residual benzene can carry over to the next reading. An exception to 
the tube replacement recommendation can be made for calibration, in that the same tube can be used 
for zeroing and subsequent span calibration, because the initial zeroing does not bring any benzene, 
other VOCs, or significant amounts of water vapor into the tube. 
 
 
 
 
 

 

 

 

Time Elapsed 

15 min. 

New 

Used 

Содержание MP186

Страница 1: ...NEO BENZ Photo ionization Detector MP186 User s Guide Version 1 01 November 2019...

Страница 2: ...ating Mode Overview 11 VOC Mode Main Menu 11 Benzene Tube Mode Main Menu 12 Benzene Tube Measurements 12 STEL Measurement 13 Benzene Tube Replacement 14 Configuration Mode 15 Navigating Config Mode 15...

Страница 3: ...he instrument After running for several minutes the residual vapor in the detector chamber will be cleared and the reading should return to zero The battery of the instrument discharges slowly even if...

Страница 4: ...een for benzene without a filtering tube and if a response over the alarm limit is detected a tube can be inserted to measure what fraction of the alarm response was actually due to benzene The 45 sec...

Страница 5: ...an alarm transducer and four keys The keys are Left Key Right Key Up Key Down Mode Key NEO BENZ user interface Alarm LEDs Left Key Move Cursor Down Mode Key On Off Right Key Move Cursor Buzzer Up Key...

Страница 6: ...s Indicates whether the pump is working or blocked Man Down Status Icon appears when Man Down alarm is enabled Lamp Status Lamp icon is shining when on and crossed out when off Bluetooth Enabled Local...

Страница 7: ...and communicate with mPower Suite WARNING To reduce the risk of ignition of hazardous atmospheres recharge and replace batteries only in areas known to be non hazardous Use only mPower s rechargeable...

Страница 8: ...sts include After self tests the instrument will show sensor information and instrument configuration information in sequence Once the startup procedure is complete and a numerical reading screen with...

Страница 9: ...ent period the gas concentration is compared with the programmed alarm limits gas concentration alarm limit settings If the concentration exceeds any of the preset limits the loud buzzer and red flash...

Страница 10: ...when datalogging is enabled In addition the following information is stored user ID site ID serial number last calibration date and alarm limits All data are retained even after the unit is turned of...

Страница 11: ...meters The Configuration Modes are slightly different depending on whether one is entering from VOC Mode or Tube Mode Some parameters are not accessible on the instrument display and the mPower Suite...

Страница 12: ...e press Down b When asked to start sampling press the Right key c Enter the temperature using the Up Down keys to increase or decrease and the Right Left keys to move the cursor Move to and press Up T...

Страница 13: ...p F 14 to 21 23 to 30 32 to 39 41 to 48 50 to 66 68 to 86 88 to 122 Run Time 360 s 240 s 170 s 110 s 60 s 45 s 30 s i Monitor the color of the tube through the holder window during the measurement If...

Страница 14: ...high VOC background and especially after a STEL measurement it is likely that the capacity of the tube to remove interfering chemicals is used up If the tube has turned from a bright orange to a pale...

Страница 15: ...t the menu item Press the Left key to save and exit Entering Numerical Values For example to enter a numerical password Increase or decrease the number from 0 through 9 by pressing the Up key or Mode...

Страница 16: ...from any of the Config Menu displays Config Mode Menus This table summarizes the Config Mode menus and sub menus Not all of these menus will be displayed if the option is not set up using mPower Suit...

Страница 17: ...17 Fig 4 Flow chart for Configuration Mode...

Страница 18: ...tor with fixed flow rate of about 0 5L min 500 cc per min to match or slightly exceed the flow rate of the instrument pump Alternatively the span gas can first be filled into a gas bag e g Tedlar bag...

Страница 19: ...enter the temperature insert a tube and press Right The message Please Apply Gas is displayed Start the gas supply and by press the Up key to begin a 30 second span calibration countdown The calibrati...

Страница 20: ...llion mg m3 milligrams per cubic meter g m3 micrograms per cubic meter mol mol mol per mol 10 6 per million Measurement Gas Measurement gases are organized in two lists Gas Library is a pre set list t...

Страница 21: ...nd TWA Alarm limits in the same manner as described above for the High Alarm limit Note All default alarm limits depend on the measurement gas and are expressed in ppm Alarm Mode There are two selecta...

Страница 22: ...er second begins Man Down Alarm Off On Press Right to make a selection followed by Left to save and exit Sensitivity This sensitivity of motion sensor can be set to Low Medium or High Low sensitivity...

Страница 23: ...nd press Up to save and exit Monitor Mnt Setup Several settings can be accessed here including date and time pump parameters display parameters and selecting 3 point calibration Date and Time Date The...

Страница 24: ...ed has its own Stall Threshold Set Pump Stall Threshold The Pump Stall Threshold is the current draw setting for a blocked pump alarm The default setting is 500 The actual real time pump current draw...

Страница 25: ...enabled the instrument re zeroes automatically if the signal drifts below the last Zero Calibration for some time Such drift can occur 1 slowly as the lamp ages 2 a dust or oil film accumulates on the...

Страница 26: ...configuration file from the instrument To switch between VOC and Benzene configuration screens change the instrument mode manually and then click Read 5 Edit the configuration parameters as desired a...

Страница 27: ...27 mPower Suite Configuration Screen for Benzene Tube Mode...

Страница 28: ...weight m w and correction factor for a 9 8 eV PID lamp The molecular weight is only needed when gas units of mg m3 are used if not leave the m w at 0 or 1 By clicking the Save User Gas box on the righ...

Страница 29: ...mple time and instantaneous reading A new Single Datalog file is created each time the instrument is turned on or the configuration is changed The middle panel shows the file start time and number of...

Страница 30: ...the same time that they appear on the PC list one can select the Rolling Graph option in the Monitor Set up configuration Select Stop when finished and Clear to discard the data when no longer needed...

Страница 31: ...or clean the existing lamp as described below 3 Use a cotton swab wetted with methanol to clean the flat window surface of the lamp If greasy dirt is hard to remove using methanol the window can be po...

Страница 32: ...should be roughly 250 350 cc min at low pump speed and 350 450 cc min at high pump speed when there is no air leakage and no tube in the holder When approaching the end of its useful operating life t...

Страница 33: ...alibration Benzene tube gasket leaks Replace Filter Clean or replace sensor module Clean or replace lamp Calibrate the unit Replace entire tube holder Buzzer inoperative Buzzer disabled Bad buzzer Che...

Страница 34: ...34 Contact Information mPower Electronics Inc 3046 Scott Blvd Santa Clara CA 95054 USA Phone 408 320 1266 Toll Free 866 mPower7 866 676 9377 Fax 669 342 7077 info mpowerinc com www mpowerinc com...

Отзывы: