Specifications
Echo/Vascular User Manual (15381-C)
Page 39 of 46
DCR-00172
Electromagnetic Compatibility
Medical electrical equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in this user manual.
Portable and mobile radio frequency (RF) communications equipment can affect
medical electrical equipment.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The product is intended for use in the electromagnetic environments specified below. The customer or the
user of the product should assure it is used in such an environment.
Emissions Test
Compliance
Guidance
RF Emissions
CISPR 11
Group 1
Group 2
The product uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment. <1>
The product must emit electromagnetic energy in order to
perform its intended function. Nearby electronic equipment may
be affected. <2>
RF Emissions
CISPR 11
Class A
The product is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic
Emissions
IEC 61000-3-2
N/A
Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3
N/A