Use Instructions
Echo/Vascular User Manual (15381-C)
Page 15 of 46
DCR-00172
Locking the Casters
The product may be equipped with total locking casters or individual locking casters.
The total locking casters have 3 functional positions: full swivel, steer, and full lock.
•
In the neutral position all four casters may swivel and roll.
•
In the steer position the front casters are held in a forward facing orientation
and the rear casters are free to swivel.
•
In the lock position all four casters are prevented from both rolling and swiveling.
Single Pedal Braking
Pedals are located on each side of the base
and are used to adjust the caster function.
Individual Locking Casters
The locking tab is located on each caster.
Push down to lock, lift to unlock.
WARNING
•
Once the product and patient have been properly positioned for the procedure, ensure the
casters are locked and the hand controller is placed in a safe position to prevent unwanted
contact and unwanted movement of the product surface during the procedure.