Specifications
Echo/Vascular User Manual (15381-C)
Page 36 of 46
DCR-00172
Environmental Conditions
ATTRIBUTE
Range for Use
Range for Storage and
Transport
Ambient Temperature
+0° to 40° C
-10° to +50° C
Relative Humidity
20% to 90% @ 30°C
–
not condensing
20% to 90% @ 30°C
–
not condensing
Atmospheric Pressure
860 to 1060 hPa
860 to 1060 hPa
Upholstery
ATTRIBUTE
SPECIFICATION
Foam
California Technical Bulletin 117
Vinyl
D.O.T. FAR 25.853 and FMVSS302
Port of NY / Boston F.D. Code
Power Requirements
ATTRIBUTE
SPECIFICATION
Electrical, Product
100
–
240 VAC, 50/60 Hz, max 5 A
Electrical, Battery Charger
100
–
240 VAC, 50/60 Hz
Battery Option
24 VDC, 2.9 Ah
Duty Cycle
10%, 2 min. on / 18 min. off
•
All electrical circuitry is isolated from chassis.
•
Grounding reliability can only be achieved when the equipment is connected to
an equivalent receptacle marked “Hospital Only” or “Hospital Grade
.
”
•
The power cord is to be used for mains disconnection.
•
Attached power supply cord set is not allowed to be diverted to other
equipment.