background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Instruction Manual 

 

 
 

SMZ Series

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Copyright © 02/02 
Motic Microscopes, European Division. 

Содержание SMZ Series

Страница 1: ...Instruction Manual SMZ Series Copyright 02 02 Motic Microscopes European Division...

Страница 2: ...Eyepiece 14 Head 18 Focus knob 20 Support collar screw 22 Incident illumination switch I 23 Transmitted illumination switch T 24 Light intensity control 2 Diopter adjuster 3 Eyepiece tubes 5 Head hold...

Страница 3: ...ents of the stereomicroscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the micr...

Страница 4: ...ent objects or for dissection For best contrast choose the side of the stage to use Warning Transmitted illumination can ONLY be used with the frosted glass stage THE HEAT GENERATED BY THE TRANSMITTED...

Страница 5: ...ntre of the stage 11 3 Turn focussing knobs 18 to mid focus range 4 The head holder 17 is mounted on a column 21 on which it can be moved up or down depending on the size of the object to be focussed...

Страница 6: ...xtended position the image is transmitted to the vertical port but the image cannot be seen by the right eyepiece A To adapt a photographic camera an adapter tube is needed This tube includes 2X lense...

Страница 7: ...hat the CS mount is responsible Place or remove it according to the procedure to obtain parfocality Maintenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL...

Страница 8: ...ination bulb a Unscrew the illuminator protector tube 7 turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the soc...

Страница 9: ...ly as possible without the head sliding down with its own weight The tension adjustment collar for focussing Fig 1 is situated between the focussing knob 18 and the head holder 17 a Loosen the screw l...

Страница 10: ...use cover badly closed Bad connection with power source Insert bulb correctly Replace bulb Close correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace wit...

Страница 11: ...ils of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through mi...

Отзывы: