B.  Electrical maintenance. 

 

1.  Changing the bulb.  

 

  Changing the transmitted illumination bulb. 

 

a.  Rest  the  stereomicroscope  on  its  side  being 

extremely  careful,  especially  with  the  eyepiece  (1) 
and the stage (11). 

 
b.  Unscrew  the  4  screws  indicated  (Fig.  3)  and  open 

the cap on the base. 

 

c.  With a cloth, carefully pull out the bulb disconnect it 

from the socket.   

 
d.  Do not touch the new bulb with your hands. Use a clean cloth to insert the pins of the 

bulb into the socket.  

 

e.  If  the  bulb  is  accidentally  touched  with  bare  hands,  it  must  be  cleaned,  as  this  could 

affect the transmission of light, and life span of the bulb.  

 

f.  Close the cover on the base and screw down firmly.  

 

  Changing incident illumination bulb. 

 

a.  Unscrew  the  illuminator  protector  tube  (7),  turning  it  anti-clockwise,  and  removing  the 

tube from that of the lenses.    

 

b.  With a cloth carefully pull out the bulb to disconnect it from the socket. 

 

g.  If  the  bulb  is  accidentally  touched  with  bare  hands,  it  must  be  cleaned,  as  this  could 

affect the transmission of light, and life span of the bulb.  

 

c.  Replace illuminator lens tube, turning in clockwise, and screwing in the protector tube.  
 

3.  Changing the fuse. 

a.  With a flat screwdriver, lightly press on the slot of the fuse holder cover (Fig. 1) and turn 

1/4 in the direction of the arrow marked.  

 

b.  Release pressure and completely remove the loosened cover. 

  

c.  Remove the fuse from the removed cover, and insert the new one, ensure that it is 0.5 

Amps.  

 

d.  Insert the cover. 

 

e.  Repeat step (a.) but turning 1/4 in the opposite direction to the arrow. The cover must be 

firmly closed.   

 
 
 
 
 

 
 
 
 
 
 

-8- 

Fig. 3 

Содержание SMZ Series

Страница 1: ...Instruction Manual SMZ Series Copyright 02 02 Motic Microscopes European Division...

Страница 2: ...Eyepiece 14 Head 18 Focus knob 20 Support collar screw 22 Incident illumination switch I 23 Transmitted illumination switch T 24 Light intensity control 2 Diopter adjuster 3 Eyepiece tubes 5 Head hold...

Страница 3: ...ents of the stereomicroscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the micr...

Страница 4: ...ent objects or for dissection For best contrast choose the side of the stage to use Warning Transmitted illumination can ONLY be used with the frosted glass stage THE HEAT GENERATED BY THE TRANSMITTED...

Страница 5: ...ntre of the stage 11 3 Turn focussing knobs 18 to mid focus range 4 The head holder 17 is mounted on a column 21 on which it can be moved up or down depending on the size of the object to be focussed...

Страница 6: ...xtended position the image is transmitted to the vertical port but the image cannot be seen by the right eyepiece A To adapt a photographic camera an adapter tube is needed This tube includes 2X lense...

Страница 7: ...hat the CS mount is responsible Place or remove it according to the procedure to obtain parfocality Maintenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL...

Страница 8: ...ination bulb a Unscrew the illuminator protector tube 7 turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the soc...

Страница 9: ...ly as possible without the head sliding down with its own weight The tension adjustment collar for focussing Fig 1 is situated between the focussing knob 18 and the head holder 17 a Loosen the screw l...

Страница 10: ...use cover badly closed Bad connection with power source Insert bulb correctly Replace bulb Close correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace wit...

Страница 11: ...ils of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through mi...

Отзывы: