Troubleshooting: 
 
ELECTRICAL PROBLEMS 

 
 

PROBLEM 

CAUSE 

SOLUTION 

The bulb does not work.  Outlet inoperative.  

 
Cable not connected.  
 
Bulb burned out. 
Fuse blown. 
Wrong Bulb. 

Have it repaired by a qualified 
specialised technician. 
Connect the cable to the power 
source.  
Replace bulb. 
Replace fuse. 
Replace with appropriate bulb. 

Bulb has short life span.  Very high voltage. 

Reduce light intensity to the minimum 
before turning on or off the  
stereomicroscope. 

Bulb burns out 
immediately. 

Wrong bulb. 

Replace with the appropriate bulb. 

Bulb flickers. 

The bulb is not inserted correctly into 
socket. 
Bulb on the point of burning out. 
Fuse cover badly closed. 
Bad connection with power source. 
 

Insert bulb correctly. 
 
Replace bulb. 
Close correctly. 
Have repaired by a qualified 
specialised technician. 

Fuse blown quickly 

Wrong fuse. 

Replace with appropriate fuse. 

Fuse blown instantly 

Short circuit. 

Have repaired by a qualified 
specialised technician. 

 
 

IMAGE QUALITY 

 
 

PROBLEM 

CAUSE 

SOLUTION 

Poor resolution. 

Eyepieces dirty. 
Objectives dirty. 

Clean eyepieces. 
Clean objectives. 

Spots, or stains in field 
of view. 

Eyepieces dirty. 

Clean eyepieces. 

* NB. Stains in field of view may also be attributed to dirt on the inside of the eyepiece. It is recommended 
therefore that the lens be cleaned by a recognised service technician.  
 
 

MECHANICAL PROBLEMS 
 
 
PROBLEM 

CAUSE 

SOLUTION 

It does not stay in focus.   The head drops down. 

Adjust the tension of the coarse focus 
knob. 

 
 
 
 
 
 
 
 
 
 
 
 
 

-11- 

Содержание SMZ Series

Страница 1: ...Instruction Manual SMZ Series Copyright 02 02 Motic Microscopes European Division...

Страница 2: ...Eyepiece 14 Head 18 Focus knob 20 Support collar screw 22 Incident illumination switch I 23 Transmitted illumination switch T 24 Light intensity control 2 Diopter adjuster 3 Eyepiece tubes 5 Head hold...

Страница 3: ...ents of the stereomicroscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the micr...

Страница 4: ...ent objects or for dissection For best contrast choose the side of the stage to use Warning Transmitted illumination can ONLY be used with the frosted glass stage THE HEAT GENERATED BY THE TRANSMITTED...

Страница 5: ...ntre of the stage 11 3 Turn focussing knobs 18 to mid focus range 4 The head holder 17 is mounted on a column 21 on which it can be moved up or down depending on the size of the object to be focussed...

Страница 6: ...xtended position the image is transmitted to the vertical port but the image cannot be seen by the right eyepiece A To adapt a photographic camera an adapter tube is needed This tube includes 2X lense...

Страница 7: ...hat the CS mount is responsible Place or remove it according to the procedure to obtain parfocality Maintenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL...

Страница 8: ...ination bulb a Unscrew the illuminator protector tube 7 turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the soc...

Страница 9: ...ly as possible without the head sliding down with its own weight The tension adjustment collar for focussing Fig 1 is situated between the focussing knob 18 and the head holder 17 a Loosen the screw l...

Страница 10: ...use cover badly closed Bad connection with power source Insert bulb correctly Replace bulb Close correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace wit...

Страница 11: ...ils of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through mi...

Отзывы: