4. MAINTENANCE & TROUBLESHOOTING

4.1 Maintenance

The biological microscope is limited to the following maintenance only:

• Turn off the power switch after use, and put on the dustproof cover after the microscope has cooled down.
• Do not operate the microscope in a room with humidity higher than 75%.
• Remove dust or ordinary dirt on optical lens surfaces with a brush, rubber suction bulb and a moistened lens tissue.
• Use only optical lens tissues and optical lens cleaner (see below). Never clean a lens with a dry optical lens tissue. 

Be sure to remove any dust before using lens tissue and cleaner.

• To remove stubborn oily or lipoid dirt (such as immersion oil or fingerprints), dip the lens tissue into a 3 to 7  

ethanol-ether mixture or a commercially available optical lens cleaning solution and then use it to wipe off the dirt.

• When cleaning an optical lens surface, wipe gently in a circle from the center to the edge.

4.2 Troubleshooting

Problem

Cause

Remedy

Can not see  
the whole field

Nosepiece is not locked into the slot

Rotate the nosepiece to lock into the slot

Condenser is not set properly

Set the condenser properly

Aperture (iris) diaphragm is not set accurately

Set the aperture (iris) diaphragm accurately

Low resolution
Poor image contrast

Incorrect opening of aperture diaphragm 

Set the opening of aperture diaphragm  
accurately

Improper focusing of condenser

Focus the condenser properly

Wrong thickness of cover slip for 0.17 
transmitted-light objective

Use the standard 0.17 thick cover slip

No immersion oil or non-specified immer-
sion oil for 100X/(oil immersion) objective

Use immersion oil supplied with the instru-
ment or go to buy cedar oil for microscope 
when supplied immersion oil used up

Bubbles in the immersion oil

Add some immersion oil or rotate the nose-
piece back and forth to remove bubbles

Immersion oil or stain left on the front lens 
of dry objective
Dirt or dust on the optical surface of objec-
tive, eyepiece, condenser or color filter

Clean the front lens of dry objective  
(see above)
Clean the dirty optical component

Poor LED  
illumination

Power plug is not plugged into the socket 
properly

Insert the power plug into the socket and 
turn on the power

LED illuminator damaged

Replace LED illuminator

Содержание RED100 Series

Страница 1: ...Instructions Manual English Motic Incorporation Ltd RED100 SERIES...

Страница 2: ......

Страница 3: ...warranty 07 2 Instrument description 09 2 1 General views 09 2 2 Part names 10 2 3 Application 12 2 4 Instrument and its major features 12 3 First time use operation 13 3 1 First time use 13 3 2 Opera...

Страница 4: ......

Страница 5: ...g to the EN 61010 1 2001 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use 1 3 Unpacking transportation storage The original shipping container a foam box in a fi...

Страница 6: ...f the protection function If there is any electrical failure of the fuse system ground protection or transformer turn off and unplug the unit immediately Make sure the microscope is set aside so it wi...

Страница 7: ...found notify the manufacturer immediately Upon receipt of the Notification of Defect as described above the manufacture is responsible to solve the problem either by repairing the defective instrumen...

Страница 8: ......

Страница 9: ...RUMENT DESCRIPTION RED 130 131 132 RED 120 Cordless model with rechargeable batteries included RED 100 101 Monocular Head Monocular Head Eyepiece Binocular Head Monocular Head Dual View Monocular Head...

Страница 10: ...head 9 Base w rechargeable batteries and external charger 3 Lock screw of eyepiece 10 Head lock screw 4 Triple nosepiece 11 Arm 5 4X 10X 40X objectives 12 Fine focus knob 6 Movable stage 13 Coarse fo...

Страница 11: ...X 40X 100X objectives 22 Coaxial coarse and fine focus knob 16 Mechanical stage 17 1 25 condenser w iris diaphragm 18 Stage adjustment knob X axis 19 Stage adjustment knob Y axis 20 Base w rechargeabl...

Страница 12: ...mpact and easy to use Separate coarse and fine focus adjustment Models RED 100 101 120 Coaxial coarse and fine focus adjustment Models RED 130 131 132 110mm 110mm gliding stage with linked slide clips...

Страница 13: ...transportation Connect the charger to the power supply Before plugging in keep in mind that the working voltage of the biological microscope shall be the same as the supply voltage Figure 1 Rotate the...

Страница 14: ...f its range Rotate the brightness control to the desired illumination Always begin examination using the low power objective Position the ob jective as close to the specimen as possible and focus by s...

Страница 15: ...ifications Figure 6 Lower the stage completely with the coarse focus knob Rotate the nosepiece to move the objective to be replaced to the side Unscrew the objective and remove it downward Fix the new...

Страница 16: ...to the slot Rotate the nosepiece to lock into the slot Condenser is not set properly Set the condenser properly Aperture iris diaphragm is not set accurately Set the aperture iris diaphragm accurately...

Страница 17: ...midity 10 40 C Below 31 C max humidity is 80 at 40 C linearly decreases to 50 Operation Permissible environment temperature Permissible relative humidity 5 40 C Below 31 C max humidity is 80 at 40 C l...

Страница 18: ...echanical tube length 160mm Mechanical stage RED 120 130 131 132 Dimension L x W Stroke L x W Position of vernier Slide clip 1 25 condenser 120x120mm 76x26mm On the right On the left of movable clip U...

Страница 19: ......

Страница 20: ...ouse 311 Gloucester Road Causeway Bay Hong Kong Tel 852 2837 0888 Fax 852 2882 2792 Motic Spain S L U Spain Pol Ind Les Corts C Les Corts 12 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax...

Отзывы: