LED REPLACEMENT

The SILVER250 Series is equipped with a 3 watt LED illumination system. The life of the LED 

may vary depending on use and intensity. To prolong the life of the LED, you should always 

turn off the unit when not in use. It is important that you only use a Motic replacement LED 

because it is integrated on to a circuit board. This LED has been tested and approved for life 

span, color temperature and brightness.

Make sure the microscope is unplugged before replacing the LED.

• Remove the eyepiece(s) from the head if they are not secured in place so they do not 

accidently fall out of the microscope. Remove the slide that may be on the stage.

• Turn the microscope on its side. Remove the 4 screws on the bottom of the microscope. 

Remove the base cover to access the LED.

• The LED is integrated on to a circuit board. This LED circuit board is held into the illumina-

tor housing by 2 screws. Remove the 2 screws from the illuminator housing to remove the 

LED circuit board.

• Unplug the LED’s power wire from the circuit board attached to the base .

• Reverse the steps listed above to install the new LED.

SILVER

250

 

SERIES

 

|  09

  

LED REPLACEMENT

SILVER

 LINE

TECHNICAL PARAMETERS

APPENDIX

Biological microscope main unit w/ monocular tube

SILVER250 Series monocular microscopes

Altitude

Transport (within package): Permissible environment temperature

Storage:
Permissible environment temperature 
Permissible relative humidity

DIMENSION (W × L × H)

 188 × 395 × 377 mm

≈ 

6.6 KG

-40 ~ +70ºC

-40 ~ +70ºC
Below 31ºC , max. humidity is 80%; 
at 40ºC , linearly decreases to 50%

Below 2000m

WEIGHT

ENVIRONMENTAL CONDITIONS

Operation:
Permissible environment temperature 
Permissible relative humidity

+5 ~ +40ºC
Below 31ºC , max. humidity is 80%; 
at 40ºC , linearly decreases to 50%

Содержание 1100104800031

Страница 1: ...SILVER250 SERIES INSTRUCTION MANUAL ENGLISH SILVERLINE...

Страница 2: ...uch the optical lens surfaces WASTE DISPOSAL Important Any damaged biological microscope must not be treated as general waste it should be disposed of according to relevant regulations OPERATION When...

Страница 3: ...ive latent infective toxic radioactive or other hazardous samples Therefore when examin ing any such sample you must comply with the relevant laws and regulations in particular the provisions related...

Страница 4: ...SERIES 04 SILVER250 SERIES The SILVER250 Series stands out for its innovation quality and student proof features offering a wide range of microscopes for educational purposes in high schools up to adv...

Страница 5: ...he minimum position before turning on or off the power switch 4 Rotate the brightness control to the desired illumination 5 After use turn the brightness control to the minimum position and then turn...

Страница 6: ...itting properties as glass Using oil slight ly increases the resolution and brightness of the image Usually a very thin slide size 1 is used for oil immersion because at this magnification the working...

Страница 7: ...ll the correct driver IMPORTANT The next time you connect the SILVER250 Series microscope to a computer make sure the USB cable is connected to the same USB port that was used during initial installat...

Страница 8: ...for 0 17 transmitted light objective No immersion oil or non specified immersion oil for 100X oil immersion objective Aperture iris diaphragm is not set accurately Incorrect opening of aperture diaphr...

Страница 9: ...access the LED The LED is integrated on to a circuit board This LED circuit board is held into the illumina tor housing by 2 screws Remove the 2 screws from the illuminator housing to remove the LED c...

Страница 10: ...tube 30 20 Length of mechanical tube Maximum field number Hinge type interpupillary distance adjustment range Observation angle Infinite 20mm 48 to 75mm 30 Triocular tube 30 20 Length of mechanical t...

Страница 11: ...epair by unauthorized parties accident alteration shipping misuse or abuse is not covered Warranty service is provided by Motic authorized technicians Determination of warranty is at the technician s...

Страница 12: ...52 2882 2792 Motic Europe S L U Spain Pol gon Industrial Les Corts C Les Corts 12 08349 Cabrera de Mar Barcelona Spain Tel 34 93 756 6286 Fax 34 93 756 6287 CCIS is a trademark of Motic Incorporation...

Страница 13: ...he appropriate immersion oil to use IMPORTANT The working distances of the 40XRD and 100XRD objectives to the slide surface are very small and although these objectives are sealed to prevent oil conta...

Страница 14: ...f the glass The finish of the microscope is hard epoxy and is resistant to acids and reagents Clean this surface with a damp cloth and mild detergent Periodically the microscope should be disassembled...

Страница 15: ...view Is the Abbe condenser in the correct position Dust or hairs seem to be moving in the image The iris diaphragm is not open wide enough Slowly open the diaphragm to increase the size of the openin...

Отзывы: