Problem Cause Remedy
The12V 20W halogen
lamp or Connect power
plug to power outlet, the
LED source does not light
up although the
microscope has been
switched on.
Power plug has not
been plugged into power
outlet.
Connect power plug to
power outlet.
12V 20W halogen lamp
or LED source is
defective.
Replace the defective 12V
20W halogen lamp or the
LED module.
The 12V 20W halogen
lamp is flickering, its light
intensity unstable, the
illumination
inhomogeneous.
End of average service
life of 12V 20W halogen
lamp has been reached.
Replace the 12V 20W
halogen lamp.
Power cable incorrectly
installed or broken.
Connect the power cable
correctly or replace it.
The pins of the 12V 20W
halogen lamp have not
been inserted correctly
into the lamp holder.
Insert the pins of the 12V
20W halogen lamp correctly
into the lamp holder.
The pins of the 12V 20W
halogen lamp have not
been inserted
symmetrically into the
lamp holder.
Insert the pins of the12V
20W halogen lamp
symmetrically into the lamp
holder.
Stage comes down by
itself, image focus is
unstable.
Stage comes down by
itself, image focus is
unstable.
Increase the torque of the
coarse focusing drive to
make motion stiffer.