4.2 Troubleshooting 

Problem Cause  Remedy 

The field of view is not 
completely visible. 

Nosepiece with objective 
has not been switched 
into click-stop position. 

Switch nosepiece with 
objective into click-stop 
position. 

Condenser has not been 
set correctly. 

Set condenser correctly. 

The aperture diaphragm 
has not been adjusted 
correctly. 

Adjust aperture 
diaphragm correctly. 

Low resolving power, poor 
image contrast. 

The aperture diaphragm 
has not been opened to 
correct size. 

Set opening of aperture 
diaphragm to correct size.

Condenser not focused 
correctly. 

Focus the condenser. 

Wrong cover glass 
thickness selected 
for use of transmitted- 
light objectives corrected 
for 0.17mm cover glass. 

Use standard 0.17 mm 
cover glass. 

Use of no or non-specified 
immersion oil with 
immersion objective. 

Use the supplied 
immersion oil. 

Air bubbles in immersion 
oil. 

Remove the bubbles by 
applying new oil or 
moving the objective to 
and fro. 

Immersion oil on the front 
lens of a dry objective. 

Clean the front lens of the 
dry objective. 

Dirt or dust on the optical 
surfaces of objectives, 
eyepieces, condenser, or 
filters 

Clean the respective 
optical components. 

Greater focus differences 
after changing the 
objective. 

The adjustable eyepieces 
have not 
been set correctly. 

Set the adjustable 
eyepieces to the 
corresponding defective 
vision. 

 

Содержание 1100104100051

Страница 1: ...ECO Series Instruction Manual WWW MOTIC COM MOTIC HONG KONG LIMITED...

Страница 2: ...mes 6 2 3 Application 7 2 4 Instrument and its major features 7 3 FIRST TIME USE AND OPERATION 8 3 1 First time use 8 3 2 Operation of the biological microscope 10 3 2 1 Interpupillary distance adjust...

Страница 3: ...ution Danger 1 2 Instrument safety The ECO microscope has been designed manufactured and inspected according to the GB 4793 1 2007 and IEC 61010 1 2001 Safety Requirements for Electrical Equipment for...

Страница 4: ...kbench for operation Since the biological microscope is a precision instrument improper operation will impair or spoil its performance The power unit is integrated in the main unit of the biological m...

Страница 5: ...vention 1 6 Quality Warranty The ECO microscope and the attached accessories are only allowed to be used for microscope examination as described in this manual The manufacturer takes no responsibility...

Страница 6: ...5 2 INSTRUMENT DESCRIPTION 2 1 General Views...

Страница 7: ...6 2 2 Parts names...

Страница 8: ...pe include Binocular siedentopf inclined at 30 degrees and rotating or Monocular siedentopf inclined at 45 degrees and rotating Widefield WF10X 18mm eyepieces Quadruple nosepiece with parcentric click...

Страница 9: ...avoid fingerprints Remove the microscope from the transport case and place it onto the worktable Retain the original packaging for storage of the instrument in longer periods of non use or for return...

Страница 10: ...ion and then turn on or off the ON OFF switch Fig 3 Fig 3 Adjust the illuminator to the desired intensity Fig 4 Fig 4 After finishing work switch off the microscope and cover it with the dust cover Th...

Страница 11: ...The adjustment of the interpupillary distance is correct when you see only one round image while looking through the two eyepieces Fig 6 Fig 6 3 2 2 Setting brightfield illumination First place a hig...

Страница 12: ...ght intensity control on the microscope stand 3 3 Converting the microscope Note Unplug the plug in power unit from line power before converting the microscope 3 3 1 Replace the 12V 20W halogen bulb o...

Страница 13: ...fingers as this will reduce bulb life If the stand is equipped with an LED source turn the focusing drive to move the mechanical stage fully up Unscrew the screws Fig 9a Take out the top part of the i...

Страница 14: ...changed into a lateral position Unscrew the objective using the rubber strip included in the delivery and remove it downward Screw the desired objective hand tight into the nosepiece as far as it will...

Страница 15: ...a brush rubber suction bulb and a moistened lens tissue Use only optical lens tissues and optical lens cleaner see below Never clean a lens with a dry optical lens tissue Be sure to remove any dust b...

Страница 16: ...ocus the condenser Wrong cover glass thickness selected for use of transmitted light objectives corrected for 0 17mm cover glass Use standard 0 17 mm cover glass Use of no or non specified immersion o...

Страница 17: ...fe of 12V 20W halogen lamp has been reached Replace the 12V 20W halogen lamp Power cable incorrectly installed or broken Connect the power cable correctly or replace it The pins of the 12V 20W halogen...

Страница 18: ...and GmbH GERMANY Christian Kremp Strasse 11 D 35578 Wetzlar Germany Tel 49 6441 210 010 Fax 49 6441 210 0122 Motic Europe SPAIN C Les Corts 12 Pol Ind Les Corts 08349 Cabrera de Mar Barcelona Spain Te...

Отзывы: